Correcting unwanted diffusion weighting in diffusion-weighted-STEAM sequence for time-dependent diffusion kurtosis imaging
Magnetic Resonance Imaging · 2025
DOI: 10.1016/j.mri.2025.110535 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2025. Verfasst von Qinfeng Zhu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, College of Biomedical Engineering & Instrument Science, Zhejiang University, Hangzhou, China. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Brain, Image Processing, Computer-Assisted, Image Interpretation, Computer-Assisted. Die Arbeit nennt 60 Literaturangaben. Gefördert von: Ministry of Science and Technology of the People's Republic of China, National Natural Science Foundation of China, Science and Technology Department of Zhejiang Province.
Abstract (Originalsprache)
PURPOSE: This study proposes an optimized acquisition protocol to minimize the unwanted diffusion weighting in DW-STEAM and facilitate time-dependent diffusion kurtosis imaging (tDKI). METHODS: We first corrected the diffusion-direction-dependent shift by optimizing the diffusion gradient amplitude. We then proposed to use low-b-value in the reference image, instead of the conventional non-diffusion-weighted (b0) acquisition, and removed the crusher gradients. The tDKI measurements from the proposed strategy were compared with conventional DWIs that included crushers in the b0 image, in a water phantom and in healthy adults (n = 8) on 3 T, and the water exchange time (τex) was calculated from the tDKI measurements. The optimal strategy was tested in five ex vivo human brains, and the results were compared with in vivo data. RESULTS: Neglecting the diffusion weighting in b0 images introduced an artificial time dependence in the apparent diffusivity of the water phantom, and resulted in elevated estimates of water exchange time (τex) in in the in vivo data. Additionally, the AIC indicated that, even when diffusion weighting in the b0 images was accounted for, the kurtosis estimation remained less stable in the in vivo than with the crusher-free approach. In contrast, using low-b-value images as reference measurements yielded reasonable tDKI estimates, with τex ranging from 15 to 40 ms for gray matter in vivo and from 20 to 60 ms for gray matter ex vivo. CONCLUSIONS: The optimized DW-STEAM acquisition eliminated artificial diffusion-time dependence, enabling the acquisition of accurate tDKI data for mapping structural morphology and transmembrane permeability.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, College of Biomedical Engineering & Instrument Science, Zhejiang University, Hangzhou, China.
- MR Collaboration, Siemens Healthcare China, Shanghai, China.
- Department of Biomedical Engineering, College of Biomedical Engineering & Instrument Science, Zhejiang University, Hangzhou, China. Electronic address: danwu.bme@zju.edu.cn.
Förderung
- Ministry of Science and Technology of the People's Republic of China
- National Natural Science Foundation of China
- Science and Technology Department of Zhejiang Province
Zitieren
APA
Zhu, Q., Ba, R., Cao, Z., Shen, Y., Li, H., Hsu, Y. C., Yan, X., & Wu, D. (2025). Correcting unwanted diffusion weighting in diffusion-weighted-STEAM sequence for time-dependent diffusion kurtosis imaging. Magnetic Resonance Imaging. https://doi.org/10.1016/j.mri.2025.110535
Vancouver
Zhu Q, Ba R, Cao Z, Shen Y, Li H, Hsu YC, et al. Correcting unwanted diffusion weighting in diffusion-weighted-STEAM sequence for time-dependent diffusion kurtosis imaging. Magnetic Resonance Imaging. 2025. doi:10.1016/j.mri.2025.110535
BibTeX
@article{Zhu2025,
author = {Zhu, Qinfeng and Ba, Ruicheng and Cao, Zuozhen and Shen, Yao and Li, Haotian and Hsu, Yi-Cheng and Yan, Xu and Wu, Dan},
title = {Correcting unwanted diffusion weighting in diffusion-weighted-STEAM sequence for time-dependent diffusion kurtosis imaging},
journal = {Magnetic Resonance Imaging},
year = {2025},
doi = {10.1016/j.mri.2025.110535}
}
EUVIMED