EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Multi‐TPS institutional surveillance of VMAT plan complexity and PSQA results over three years

Sotiri Stathakis, Ara Alexandrian, Mason Heath

Journal of Applied Clinical Medical Physics · 2026-07 · 27 (7)

DOI: 10.1002/acm2.70691 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 06.07.2026. Verfasst von Sotiri Stathakis und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics and Radiation Oncology, Mary Bird Perkins Cancer Center, Baton Rouge, Louisiana, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Quality Assurance, Health Care, Neoplasms. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Running VMAT plan‐complexity surveillance directly inside an oncology information system (OIS) makes it practical to monitor plan quality across a large, multi‐linac program. Complexity metrics are increasingly used to anticipate patient‐specific QA (PSQA) outcomes, but few studies span multiple years or tie complexity monitoring to the OIS at scale, which leaves published benchmarks hard to generalize and prospective monitoring slow to enter routine practice.

Purpose

To describe an automated Python pipeline integrated with the MOSAIQ OIS for computing eight established VMAT plan complexity metrics from DICOM‐RT files at scale, and to characterize the resulting distributions across a three‐year institutional VMAT program stratified by treatment planning system (TPS), anatomic site, dosimetrist, and calendar year, with correlation to matched PSQA outcomes.

Methods

All clinically approved VMAT plans from March 2023 to March 2026 were identified via MOSAIQ OIS (v2.6, Elekta AB) at a multisite comprehensive cancer center. A purpose‐built Python tool queried the MOSAIQ SQL Server database and parsed DICOM‐RT plan files to compute MU Factor, mean aperture area (MAA), mean leaf gap (MLG), aperture irregularity (AI), small aperture scores (SASs) at 5 and 10 mm, modulation complexity score for VMAT (MCSv), and Modulation Index Total (MITotal). Plans were stratified by TPS (Pinnacle3 vs. Monaco), anatomic site group (11 groups), dosimetrist, and year. Mann–Whitney U and Kruskal–Wallis tests were applied for group comparisons. PSQA pass rates (3%/2 mm global gamma, ≥95% threshold) were extracted from MOSAIQ and matched to complexity data for 3402 of 3751 plans (90.7%); Spearman rank correlations quantified complexity–pass‐rate associations.

Results

The institutional median (IQR) MU Factor was 2.96 (2.26–4.00) MU/cGy and MCSv was 0.966 (0.955–0.978). Monaco plans showed significantly higher MU Factors (+32%, p < 0.001), narrower apertures (MLG −16%), and higher MCSv (+1.0%) than Pinnacle3 plans. Pelvis and head and neck plans had the highest median MU Factors (3.77 and 3.40 MU/cGy); Breast plans were most efficient (2.10 MU/cGy). Institutional MU Factor rose 20% over three years. Of 3402 PSQA‐matched plans, the mean gamma pass rate was 98.7% (median 99.6%); 115 (3.4%) fell below 95%. All complexity metrics correlated significantly with pass rate (all p < 0.001); AI (ρ = −0.358) and MCSv (ρ = +0.336) were the strongest predictors. Monaco plans achieved a higher mean PSQA pass rate than Pinnacle3 (99.0% vs. 98.5–98.6%) despite greater modulation. Significant linac‐level variation in pass rates was observed (97.3%–99.2%), not fully explained by plan complexity.

Conclusions

OIS‐integrated VMAT complexity and PSQA surveillance is feasible at scale and yields clinically useful information. AI and MCSv were the strongest complexity‐based predictors of PSQA outcome. The TPS‐, site‐, and linac‐specific complexity profiles we observed argue for risk‐stratified PSQA design. The pipeline is readily transferable to other MOSAIQ‐based programs for prospective plan‐quality monitoring.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedRadiotherapy, Intensity-ModulatedQuality Assurance, Health CareNeoplasmsHumansRadiotherapy DosageOrgans at RiskAlgorithmsParticle Accelerators

Stichwörter der Autor:innen

MCSvMOSAIQMU factorMonacoPinnacleVMATdosimetrist variabilityoncology information systemplan complexityquality assurancetreatment planning system

Autor:innen und Institutionen

  • Department of Medical Physics and Radiation Oncology, Mary Bird Perkins Cancer Center, Baton Rouge, Louisiana, USA.

Zitieren

APA

Stathakis, S., Alexandrian, A., & Heath, M. (2026). Multi‐TPS institutional surveillance of VMAT plan complexity and PSQA results over three years. Journal of Applied Clinical Medical Physics, 27(7). https://doi.org/10.1002/acm2.70691

Vancouver

Stathakis S, Alexandrian A, Heath M. Multi‐TPS institutional surveillance of VMAT plan complexity and PSQA results over three years. Journal of Applied Clinical Medical Physics. 2026;27(7). doi:10.1002/acm2.70691

BibTeX
@article{Stathakis2026,
  author = {Stathakis, Sotiri and Alexandrian, Ara and Heath, Mason},
  title = {Multi‐TPS institutional surveillance of VMAT plan complexity and PSQA results over three years},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {7},
  doi = {10.1002/acm2.70691}
}
Kostenlos starten →