Commissioning and clinical implementation of an independent dose calculation system for scanned proton beams
Journal of Applied Clinical Medical Physics · 2024-05 · 25 (5)
DOI: 10.1002/acm2.14328 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 29.03.2024. Verfasst von Ralf Dreindl und 9 weiteren Autor:innen. Beteiligte Einrichtung: MedAustron Ion Therapy Center, Wiener Neustadt, Austria. Hauptthemen laut MEDLINE: Proton Therapy, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted, Software. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Experimental patient‐specific QA (PSQA) is a time and resource‐intensive process, with a poor sensitivity in detecting errors. Radiation therapy facilities aim to substitute it by means of independent dose calculation (IDC) in combination with a comprehensive beam delivery QA program. This paper reports on the commissioning of the IDC software tool myQA iON (IBA Dosimetry) for proton therapy and its clinical implementation at the MedAustron Ion Therapy Center.
Methods
The IDC commissioning work included the validation of the beam model, the implementation and validation of clinical CT protocols, and the evaluation of patient treatment data. Dose difference maps, gamma index distributions, and pass rates (GPR) have been reviewed. The performance of the IDC tool has been assessed and clinical workflows, simulation settings, and GPR tolerances have been defined.
Results
Beam model validation showed agreement of ranges within ± 0.2 mm, Bragg‐Peak widths within ± 0.1 mm, and spot sizes at various air gaps within ± 5% compared to physical measurements. Simulated dose in 2D reference fields deviated by ‐0.3% ± 0.5%, while 3D dose distributions differed by 1.8% on average to measurements. Validation of the CT calibration resulted in systematic differences of 2.0% between IDC and experimental data for tissue like samples. GPRs of 99.4 ± 0.6% were found for head, head and neck, and pediatric CT protocols on a 2%/2 mm gamma criterion. GPRs for the adult abdomen protocol were at 98.9% on average with 3%/3 mm. Root causes of GPR outliers, for example, implants were identified and evaluated.
Conclusion
IDC has been successfully commissioned and integrated into the MedAustron clinical workflow for protons in 2021. IDC has been stepwise and safely substituting experimental PSQA since February 2021. The initial reduction of proton experimental PSQA was about 25% and reached up to 90% after 1 year.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- MedAustron Ion Therapy Center, Wiener Neustadt, Austria.
- Department of Medical Physics and Radiation Protection, Hospital Universitario La Princesa, Madrid, Spain.
- Division Medical Radiation Physics, Department of Radiation Oncology, Medical University of Vienna/AKH Wien, Vienna, Austria.
- Department of Isotope Physics, Faculty of Physics, University of Vienna, Vienna, Austria.
- Medical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
ORCID-Profile: Ralf Dreindl
Zitieren
APA
Dreindl, R., Bolsa-Ferruz, M., Fayos-Sola, R., Padilla Cabal, F., Scheuchenpflug, L., Elia, A., Amico, A., Carlino, A., Stock, M., & Grevillot, L. (2024). Commissioning and clinical implementation of an independent dose calculation system for scanned proton beams. Journal of Applied Clinical Medical Physics, 25(5). https://doi.org/10.1002/acm2.14328
Vancouver
Dreindl R, Bolsa-Ferruz M, Fayos-Sola R, Padilla Cabal F, Scheuchenpflug L, Elia A, et al. Commissioning and clinical implementation of an independent dose calculation system for scanned proton beams. Journal of Applied Clinical Medical Physics. 2024;25(5). doi:10.1002/acm2.14328
BibTeX
@article{Dreindl2024,
author = {Dreindl, Ralf and Bolsa-Ferruz, Marta and Fayos-Sola, Rosa and Padilla Cabal, Fatima and Scheuchenpflug, Lukas and Elia, Alessio and Amico, Antonio and Carlino, Antonio and Stock, Markus and Grevillot, Loïc},
title = {Commissioning and clinical implementation of an independent dose calculation system for scanned proton beams},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
volume = {25},
number = {5},
doi = {10.1002/acm2.14328}
}