Error detection using a multi‐channel hybrid network with a low‐resolution detector in patient‐specific quality assurance
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14327 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Bing Yan und 6 weiteren Autor:innen. Beteiligte Einrichtung: School of Instrument Science and Optoelectronics Engineering, Hefei University of Technology, Hefei, China. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China.
Abstract (Originalsprache)
Purpose
This study aimed to develop a hybrid multi‐channel network to detect multileaf collimator (MLC) positional errors using dose difference (DD) maps and gamma maps generated from low‐resolution detectors in patient‐specific quality assurance (QA) for Intensity Modulated Radiation Therapy (IMRT).
Methods
A total of 68 plans with 358 beams of IMRT were included in this study. The MLC leaf positions of all control points in the original IMRT plans were modified to simulate four types of errors: shift error, opening error, closing error, and random error. These modified plans were imported into the treatment planning system (TPS) to calculate the predicted dose, while the PTW seven29 phantom was utilized to obtain the measured dose distributions. Based on the measured and predicted dose, DD maps and gamma maps, both with and without errors, were generated, resulting in a dataset with 3222 samples. The network's performance was evaluated using various metrics, including accuracy, sensitivity, specificity, precision, F1‐score, ROC curves, and normalized confusion matrix. Besides, other baseline methods, such as single‐channel hybrid network, ResNet‐18, and Swin‐Transformer, were also evaluated as a comparison.
Results
The experimental results showed that the multi‐channel hybrid network outperformed other methods, demonstrating higher average precision, accuracy, sensitivity, specificity, and F1‐scores, with values of 0.87, 0.89, 0.85, 0.97, and 0.85, respectively. The multi‐channel hybrid network also achieved higher AUC values in the random errors (0.964) and the error‐free (0.946) categories. Although the average accuracy of the multi‐channel hybrid network was only marginally better than that of ResNet‐18 and Swin Transformer, it significantly outperformed them regarding precision in the error‐free category.
Conclusion
The proposed multi‐channel hybrid network exhibits a high level of accuracy in identifying MLC errors using low‐resolution detectors. The method offers an effective and reliable solution for promoting quality and safety of IMRT QA.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Instrument Science and Optoelectronics Engineering, Hefei University of Technology, Hefei, China.
- Department of Radiation Oncology, The First Affiliated Hospital of University of Science and Technology of China, Hefei, China.
- School of Computer Science and Technology, University of Science and Technology of China, Hefei, China.
- Department of Radiation Oncology, Hubei Cancer Hospital, TongJi Medical College, Huazhong University of Science and Technology, Wuhan, China.
- Department of Radiation Oncology, Rutgers-Cancer Institute of New Jersey, Rutgers-Robert Wood Johnson Medical School, New Brunswick, New Jersey, USA.
ORCID-Profile: Xiao Wang
Förderung
- National Natural Science Foundation of China
Zitieren
APA
Yan, B., Shi, J., Xue, X., Peng, H., Wu, A., Wang, X., & Ma, C. (2024). Error detection using a multi‐channel hybrid network with a low‐resolution detector in patient‐specific quality assurance. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14327
Vancouver
Yan B, Shi J, Xue X, Peng H, Wu A, Wang X, et al. Error detection using a multi‐channel hybrid network with a low‐resolution detector in patient‐specific quality assurance. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14327
BibTeX
@article{Yan2024,
author = {Yan, Bing and Shi, Jun and Xue, Xudong and Peng, Hu and Wu, Aidong and Wang, Xiao and Ma, Chi},
title = {Error detection using a multi‐channel hybrid network with a low‐resolution detector in patient‐specific quality assurance},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14327}
}
EUVIMED