Effect of modulation factor and low dose threshold level on gamma pass rates of single isocenter multi‐target SRT treatment plans

Elena Timakova, Sergei F Zavgorodni

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14459 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Elena Timakova und 1 weiteren Autor:innen. Beteiligte Einrichtung: University of Victoria, Victoria, British Columbia, Canada. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Radiotherapy, Intensity-Modulated, Algorithms. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 59 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

SRS MapCHECK (SMC) is a commercially available patient‐specific quality assurance (PSQA) tool for stereotactic radiosurgery (SRS) applications. This study investigates the effects of degree of modulation, location off‐axis, and low dose threshold (LDT) selection on gamma pass rates (GPRs) between SMC and treatment planning system, Analytical Anisotropic Algorithm (AAA), or Vancouver Island Monte Carlo (VMC++ algorithm) system calculated dose distributions.

Methods

Volumetric‐modulated arc therapy (VMAT) plans with modulation factors (MFs) ranging from 2.7 to 10.2 MU/cGy were delivered to SMC at isocenter and 6 cm off‐axis. SMC measured dose distributions were compared against AAA and VMC++ via gamma analysis (3%/1 mm) with LDT of 10% to 80% using SNC Patient software.

Results

Comparing on‐axis SMC dose against AAA and VMC++ with LDT of 10%, all AAA‐calculated plans met the acceptance criteria of GPR ≥ 90%, and only one VMC++ calculated plan was marginally outside the acceptance criteria with pass rate of 89.1%. Using LDT of 80% revealed decreasing GPR with increasing MF. For AAA, GPRs reduced from 100% at MF of 2.7 MU/cGy to 57% at MF of 10.2 MU/cGy, and for VMC++ calculated plans, the GPRs reduced from 89% to 60% in the same MF range. Comparison of SMC dose off‐axis against AAA and VMC++ showed more pronounced reduction of GPR with increasing MF. For LDT of 10%, AAA GPRs reduced from 100% to 83% in the MF range of 2.7 to 9.8 MU/cGy, and VMC++ GPR reduced from 100% to 91% in the same range. With 80% LDT, GPRs dropped from 100% to 42% for both algorithms.

Conclusions

MF, dose calculation algorithm, and LDT selections are vital in VMAT‐based SRT PSQA. LDT of 80% enhances sensitivity of gamma analysis for detecting dose differences compared to 10% LDT. To achieve better agreement between calculated and SMC dose, it is recommended to limit the MF to 4.6 MU/cGy on‐axis and 3.6 MU/cGy off‐axis.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedRadiotherapy DosageRadiotherapy, Intensity-ModulatedAlgorithmsRadiosurgeryMonte Carlo MethodHumansOrgans at RiskQuality Assurance, Health CareSoftwarePhantoms, ImagingNeoplasms

Stichwörter der Autor:innen

AAAGPRMonte CarloSIMT SRTSRS MapCHECKVMATVMC++dosimetrygamma analysismodulation factorpatient specific QAplan complexityverification

Autor:innen und Institutionen

  • University of Victoria, Victoria, British Columbia, Canada.
  • BC Cancer Agency, Vancouver Island Centre, Victoria, British Columbia, Canada.

Zitieren

APA

Timakova, E., & Zavgorodni, S. F. (2024). Effect of modulation factor and low dose threshold level on gamma pass rates of single isocenter multi‐target SRT treatment plans. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14459

Vancouver

Timakova E, Zavgorodni SF. Effect of modulation factor and low dose threshold level on gamma pass rates of single isocenter multi‐target SRT treatment plans. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14459

BibTeX
@article{Timakova2024,
  author = {Timakova, Elena and Zavgorodni, Sergei F},
  title = {Effect of modulation factor and low dose threshold level on gamma pass rates of single isocenter multi‐target SRT treatment plans},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14459}
}
Kostenlos starten →