EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

AFI manual planning versus HyperArc auto‐planning: A head‐to‐head comparison of SRS plan quality

Dharmin D Desai, Ivan L Cordrey, E Lee Johnson, Thomas A Oldland

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14503 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Dharmin D Desai und 3 weiteren Autor:innen. Beteiligte Einrichtung: Varian Medical Systems Inc., Advanced Oncology Solutions, Hixson, Tennessee, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Radiotherapy, Intensity-Modulated, Radiosurgery. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

HyperArc (HA) auto‐planning offers simplicity for the end user and consistently high‐quality SRS plans. The “Ask For It” (AFI) optimization strategy offers a manual planning technique that, when coupled with R50% Analytic , can be guided to deliver a plan with an intermediate dose spill “as low as reasonably achievable” and high target dose conformity. A direct comparison of SRS plan quality obtained using the manual planning AFI strategy and HA has been performed.

Methods

Using a CT data set available from the Radiosurgery Society, 54 PTVs were created and used to generate 19 individual SRS/SRT cases. Case complexity ranged from single PTV plans to multiple PTV plans with a single isocenter. PTV locations ranged from relative isolation from critical structures to lesions within 1.5 mm of the optic apparatus and abutting the brainstem. All cases were planned using both the AFI and HA optimization strategies as implemented in the Varian Medical Systems Eclipse Treatment Planning System. A range of treatment plan quality metrics were obtained including Intermediate Dose Spill (R50%), Conformity Indices CI RTOG and CI Paddick , PTV Dose Coverage (Dn%), PTV Mean Dose, and Modulation Factor. The Wilcoxon Signed Rank Sum non‐parametric statistical method was utilized to compare the obtained plan quality metrics.

Results

Statistically significant improvements were found for the AFI strategy for metrics R50%, CI RTOG , CI Paddick , and PTV Mean Dose ( p < 0.001). HA achieved superior coverage for Dn% ( p = 0.018), while the Modulation Factors were not significantly different for AFI compared to HA optimization ( p = 0.13).

Conclusion

This study provides evidence that the AFI manual planning strategy can produce high‐quality planning metrics similar to the HA auto‐planning method.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedRadiotherapy DosageRadiotherapy, Intensity-ModulatedRadiosurgeryOrgans at RiskHumansAlgorithmsQuality Assurance, Health CareBrain NeoplasmsNeoplasms

Stichwörter der Autor:innen

AFIHyperArcR50%SIMTSRSplan quality

Autor:innen und Institutionen

  • Varian Medical Systems Inc., Advanced Oncology Solutions, Hixson, Tennessee, USA.
  • Thompson Cancer Survival Center, Cumberland Medical Center, Crossville, Tennessee, USA.
  • Department of Radiation Medicine, University of Kentucky Chandler Medical Center, Lexington, Kentucky, USA.

Förderung

  • NCI NIH HHS

Zitieren

APA

Desai, D. D., Cordrey, I. L., Johnson, E. L., & Oldland, T. A. (2024). AFI manual planning versus HyperArc auto‐planning: A head‐to‐head comparison of SRS plan quality. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14503

Vancouver

Desai DD, Cordrey IL, Johnson EL, Oldland TA. AFI manual planning versus HyperArc auto‐planning: A head‐to‐head comparison of SRS plan quality. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14503

BibTeX
@article{Desai2024,
  author = {Desai, Dharmin D and Cordrey, Ivan L and Johnson, E Lee and Oldland, Thomas A},
  title = {AFI manual planning versus HyperArc auto‐planning: A head‐to‐head comparison of SRS plan quality},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14503}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche