Validation of MLC leaf open time calculation methods for PSQA in adaptive radiotherapy with tomotherapy units
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14478 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Marie Nasrallah und 6 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Radiation Physics, Lausanne University Hospital and Lausanne University, Lausanne, Switzerland. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Radiotherapy Dosage, Phantoms, Imaging. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Accuray Inc..
Abstract (Originalsprache)
Background
Treatment delivery safety and accuracy are essential to control the disease and protect healthy tissues in radiation therapy. For usual treatment, a phantom‐based patient specific quality assurance (PSQA) is performed to verify the delivery prior to the treatment. The emergence of adaptive radiation therapy (ART) adds new complexities to PSQA. In fact, organ at risks and target volume re‐contouring as well as plan re‐optimization and treatment delivery are performed with the patient immobilized on the treatment couch, making phantom‐based pretreatment PSQA impractical. In this case, phantomless PSQA tools based on multileaf collimator (MLC) leaf open times (LOTs) verifications provide alternative approaches for the Radixact® treatment units. However, their validity is compromised by the lack of independent and reliable methods for calculating the LOT performed by the MLC during deliveries.
Purpose
To provide independent and reliable methods of LOT calculation for the Radixact® treatment units.
Methods
Two methods for calculating the LOTs performed by the MLC during deliveries have been implemented. The first method uses the signal recorded by the build‐in detector and the second method uses the signal recorded by optical sensors mounted on the MLC. To calibrate the methods to the ground truth, in‐phantom ionization chamber LOT measurements have been conducted on a Radixact® treatment unit. The methods were validated by comparing LOT calculations with in‐phantom ionization chamber LOT measurements performed on two Radixact® treatment units.
Results
The study shows a good agreement between the two LOT calculation methods and the in‐phantom ionization chamber measurements. There are no notable differences between the two methods and the same results were observed on the different treatment units.
Conclusions
The two implemented methods have the potential to be part of a PSQA solution for ART in tomotherapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Radiation Physics, Lausanne University Hospital and Lausanne University, Lausanne, Switzerland.
- Accuray Incorporated, Madison, Wisconsin, USA.
- Radiation Oncology Department, Lausanne University Hospital and Lausanne University, Lausanne, Switzerland.
ORCID-Profile: Raphaël Moeckli
Förderung
- Accuray Inc.
Zitieren
APA
Nasrallah, M., Bochud, F., Tellapragada, N., Bourhis, J., Chao, E., Casey, D., & Moeckli, R. (2024). Validation of MLC leaf open time calculation methods for PSQA in adaptive radiotherapy with tomotherapy units. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14478
Vancouver
Nasrallah M, Bochud F, Tellapragada N, Bourhis J, Chao E, Casey D, et al. Validation of MLC leaf open time calculation methods for PSQA in adaptive radiotherapy with tomotherapy units. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14478
BibTeX
@article{Nasrallah2024,
author = {Nasrallah, Marie and Bochud, François and Tellapragada, Neelima and Bourhis, Jean and Chao, Edward and Casey, Dylan and Moeckli, Raphaël},
title = {Validation of MLC leaf open time calculation methods for PSQA in adaptive radiotherapy with tomotherapy units},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14478}
}
EUVIMED