Characterization of a novel plastic scintillation detector for in vivo electron dosimetry
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70623 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Cornelius J Bauer und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University Medical Centre Mannheim, Heidelberg University, Mannheim, Germany. Hauptthemen laut MEDLINE: Electrons, Scintillation Counting, Plastics, Phantoms, Imaging. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Real‐time dosimetry of surface doses in electron beams has not been widely established yet. Plastic scintillation detectors (PSDs) promise high spatial resolution and real‐time dosimetry with minimum perturbation of the radiation field. This study characterizes a novel PSD in an electron beam to determine its suitability for in vivo dosimetry.
Methods
Dual‐channel Cherenkov radiation correction and dosimetric characterization of the PSD were investigated using reference ionization chambers. Percentage depth‐dose curves, lateral profiles, and output factors were compared with reference ionization chamber measurements. Surface doses were measured on solid water and on an anthropomorphic phantom and were compared to ionization chamber and radiochromic film measurements.
Results
The investigated PSD demonstrated clinically acceptable linearity, dose rate independence, isotropy and reproducibility (total variation < 2%). Dosimetric deviations in R 50 and R 80 were below 1.0 mm and lateral profiles agreed with a mean absolute error below 1.5%. Small field measurements were within 2% of the reference ionization chamber results. Surface dose measurements had mean relative deviations of 1.3% from ionization chamber measurements and 2.1% from radiochromic film measurements.
Conclusion
The PSD investigated in this study is suitable for clinically acceptable electron beam dosimetry and provides accurate dosimetric results for surface dose measurements. It has the potential to be used for real‐time in vivo dosimetry.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiation Oncology, University Medical Centre Mannheim, Heidelberg University, Mannheim, Germany.
- DKFZ-Hector Cancer Institute, University Medical Centre Mannheim, Mannheim, Germany.
- Mannheim Institute for Intelligent Systems in Medicine (MIISM), Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
ORCID-Profile: Cornelius J Bauer, Jens Fleckenstein
Zitieren
APA
Bauer, C. J., Schneider, F., Göbel, I. D., Oppitz, H., Giordano, F. A., & Fleckenstein, J. (2026). Characterization of a novel plastic scintillation detector for in vivo electron dosimetry. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70623
Vancouver
Bauer CJ, Schneider F, Göbel ID, Oppitz H, Giordano FA, Fleckenstein J. Characterization of a novel plastic scintillation detector for in vivo electron dosimetry. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70623
BibTeX
@article{Bauer2026,
author = {Bauer, Cornelius J and Schneider, Frank and Göbel, Ida D and Oppitz, Hans and Giordano, Frank A and Fleckenstein, Jens},
title = {Characterization of a novel plastic scintillation detector for in vivo electron dosimetry},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70623}
}
EUVIMED