Sensitivity and concordance study of an open source EPID‐based linear accelerator test suite
Journal of Applied Clinical Medical Physics · 2026-04 · 27 (4)
DOI: 10.1002/acm2.70569 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 14.04.2026. Verfasst von Lana C Critchfield und 14 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Michigan, Wyoming, Michigan, USA. Hauptthemen laut MEDLINE: Particle Accelerators, Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, Varian Medical Systems.
Abstract (Originalsprache)
Background
Electronic portal imaging devices (EPIDs) enhance linear accelerator (linac) quality assurance (QA) efficiency but are not universally used beyond patient‐specific IMRT QA. A multi‐institutional Consortium seeks to standardize linac QA utilizing an EPID‐based test suite.
Purpose
This study evaluates the sensitivity of a test suite to intentional linac and plan errors and the concordance with linac fluctuations over time.
Methods
Baseline and error‐introduced tests were performed on a Varian TrueBeam and Clinac. Evaluated parameters included enhanced dynamic wedge factors (EDW) (6MV), central axis dose (6MV, 6MV FFF, 16MV), focal spot alignment and beam symmetry (6MV, 16MV), dose rate‐gantry speed and multi‐leaf collimator (MLC) leaf speed (6MV). Errors included varying EDW angles (10°–60°), output changes, beam steering for focal spot and symmetry inaccuracies, and meterset adjustments of control points for dose rate‐gantry speed and MLC leaf speed tests. Measurements were performed with an EPID, analyzed via test suite software, and compared to an ion chamber (IC) or array. Concordance analysis was performed on a single linac over 5 months to assess the test suite's ability to quantify deviations during routine operations. Measurement reproducibility was assessed.
Results
Results were reported as differences between EPID and an IC or array. For each EDW, wedge factor differences were
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, University of Michigan, Wyoming, Michigan, USA.
- Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan, USA.
- Department of Radiation Oncology, University of Michigan, Essexville, Michigan, USA.
- Department of Radiation Oncology, Washington University, St. Louis, Missouri, USA.
- Department of Radiation Oncology, Calvary Mater Hospital Newcastle, NSW, Australia.
- Memorial Sloan Kettering Cancer Center, Samford, Connecticut, USA.
ORCID-Profile: Michael Barnes, Seng Boh Lim, Nilendu Gupta, Richard A Popple, Jean M Moran
Förderung
- NCI NIH HHS
- Varian Medical Systems
Zitieren
APA
Critchfield, L. C., Knill, C. S., Durussel, E., Mikell, J. K., Barnes, M., Pillainayagam, M., Donahue, W. P., Lim, S. B., DeMarco, J. J., Perez, M., Gupta, N., Al-Hallaq, H., Popple, R. A., Irrer, J., & Moran, J. M. (2026). Sensitivity and concordance study of an open source EPID‐based linear accelerator test suite. Journal of Applied Clinical Medical Physics, 27(4). https://doi.org/10.1002/acm2.70569
Vancouver
Critchfield LC, Knill CS, Durussel E, Mikell JK, Barnes M, Pillainayagam M, et al. Sensitivity and concordance study of an open source EPID‐based linear accelerator test suite. Journal of Applied Clinical Medical Physics. 2026;27(4). doi:10.1002/acm2.70569
BibTeX
@article{Critchfield2026,
author = {Critchfield, Lana C and Knill, Cory S and Durussel, Ellie and Mikell, Justin K and Barnes, Michael and Pillainayagam, Michael and Donahue, William P and Lim, Seng Boh and DeMarco, John J and Perez, Mario and Gupta, Nilendu and Al-Hallaq, Hania and Popple, Richard A and Irrer, Jim and Moran, Jean M},
title = {Sensitivity and concordance study of an open source EPID‐based linear accelerator test suite},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {4},
doi = {10.1002/acm2.70569}
}
EUVIMED