Evaluation of QUASAR Insight Phantom for daily imaging QA of MRgRT linacs

Eli Grant, Jiayi Liu, Ronak Etemadpour, Chengyu Shi, Huisi Ai, Percy Lee, Terence Williams, An Liu, Borna Maraghechi

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70463 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Eli Grant und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, City of Hope National Medical Center, Irvine, California, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Quality Assurance, Health Care, Particle Accelerators. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Magnetic resonance imaging (MRI) and its importance in modern radiation therapy requires a standardized daily quality assurance (QA) procedure that is both comprehensive and efficient. Current imaging QA guidelines require a more inclusive analysis of imaging characteristics, as well as a simplification of setup procedures. We have studied the feasibility of using the Insight Phantom to provide consistent images for analyses of uniformity, slice thickness, spatial resolution, and distortion across a large field of view (FOV).

Methods

The large imaging region of the Insight Phantom was imaged using 3D True Fast Imaging with Steady State Precession (TruFi) and analyzed using the Modus QA software. The base and upright section were imaged in coronal, axial, and sagittal orientations over 3 months.

Results

Geometric distortion analysis yielded average maximum distortions within a 210 mm diameter of 0.57, 0.73, and 0.45 mm for axial, sagittal, and coronal orientations, respectively. Similarly, within a 310 mm diameter, average maximum distortions were 0.58, 0.73, and 0.74 mm with no distortions greater than or equal to 1 mm. Average modulation transfer function (MTF) values were 0.431 ± 0.007, 0.436 ± 0.007, and 0.433 ± 0.010 lp/mm for axial, sagittal, and coronal images, respectively. Finally, slice thickness values were consistent at 2.19 ± 0.04, 2.22 ± 0.06, and 2.24 ± 0.07 mm for axial, sagittal, and coronal imaging orientations, respectively. Images were repeatable and setup procedures were quick and straightforward, enhancing therapist workflow and efficiency.

Conclusion

The Insight Phantom's large coverage allows a more in‐depth analysis of the MRI's large FOV, and its simple setup ensures a repeatable daily QA procedure for therapists. The software also provides a comprehensive analysis of key imaging parameters in a detailed report that can be used to monitor MRI functionality and quality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingMagnetic Resonance ImagingQuality Assurance, Health CareParticle AcceleratorsRadiotherapy Planning, Computer-AssistedImage Processing, Computer-AssistedHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedSoftware

Stichwörter der Autor:innen

MR imagingadaptive radiation therapylinacmagnetic resonance‐guided radiation therapyphantomquality assurance

Autor:innen und Institutionen

  • Department of Radiation Oncology, City of Hope National Medical Center, Irvine, California, USA.
  • Department of Radiation Oncology, University of California Irvine Medical Center, Orange, California, USA.
  • Department of Radiation Oncology, City of Hope National Medical Center, Duarte, California, USA.

Zitieren

APA

Grant, E., Liu, J., Etemadpour, R., Shi, C., Ai, H., Lee, P., Williams, T., Liu, A., & Maraghechi, B. (2026). Evaluation of QUASAR Insight Phantom for daily imaging QA of MRgRT linacs. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70463

Vancouver

Grant E, Liu J, Etemadpour R, Shi C, Ai H, Lee P, et al. Evaluation of QUASAR Insight Phantom for daily imaging QA of MRgRT linacs. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70463

BibTeX
@article{Grant2026,
  author = {Grant, Eli and Liu, Jiayi and Etemadpour, Ronak and Shi, Chengyu and Ai, Huisi and Lee, Percy and Williams, Terence and Liu, An and Maraghechi, Borna},
  title = {Evaluation of QUASAR Insight Phantom for daily imaging QA of MRgRT linacs},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70463}
}
Kostenlos starten →