Investigation of an EPID‐based dose verification system for patient‐specific quality assurance in a harmony device

Dohyeon Yoo iD, Hyunggun Lee, Sangmin Lee, Soorim Han, Taeho Kim, Changhwan Kim iD, Min Cheol Han iD, Jin Sung Kim

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70201 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Dohyeon Yoo und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Yonsei Cancer Center, Yonsei University, College of Medicine & Heavy Ion Therapy Research Institute, Yonsei Univers…. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Particle Accelerators, Quality Assurance, Health Care, Radiotherapy, Intensity-Modulated. Die Arbeit nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Precise patient‐specific quality assurance (PSQA) is critically important in advanced treatments.

Purpose

This study aimed to evaluate the performance of the RadCalc software's electronic portal imaging device (EPID)‐based patient specific QA function, focusing on its ability to accurately reconstruct doses from EPID images acquired on the Elekta Harmony Pro linear accelerators (LINACs).

Methods

Beam modeling was conducted for 6‐MV and 6‐MV flattening filter–free (FFF) photon beams on the recently released Harmony Pro LINAC system in RadCalc. Volumetric modulated arc therapy (VMAT) plans for 6‐MV (19 patients) and 6‐MV FFF (10 patients) were generated using the ArcCHECK phantom to replicate clinical treatments. Dose calculations from RadCalc, using portal images, were compared to those generated by the treatment planning system (TPS) through gamma analysis at 1%/1, 2%/2, and 3%/3 mm, with a 90% minimum passing rate.

Results

For 6‐MV and 6‐MV FFF beams, gamma analysis comparing EPID data with RadCalc showed a 98.83% passing rate at 2%/2 mm. For VMAT plans, 6‐MV beams achieved the highest passing rate (99.88%) for the breast case at 3%/3 mm and the lowest (82.33%) for the lymphatic node case. Similarly, 6‐MV FFF beams yielded the highest passing rate (99.87%) for the abdominal case at 3%/3 mm and the lowest (94.07%) for the bone case.

Conclusions

RadCalc showed high accuracy for dose evaluation on the Elekta Harmony Pro LINAC. Beam modeling achieved a gamma passing rate of at least 98% (2%/2 mm), demonstrating reliable performance even under minor setup variations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedParticle AcceleratorsQuality Assurance, Health CareRadiotherapy, Intensity-ModulatedPhantoms, ImagingSoftwareNeoplasmsElectrical Equipment and SuppliesRadiometryHumansRadiotherapy DosagePhotons

Stichwörter der Autor:innen

IMRTRadCalcVMATharmony propatient‐specific quality assurance

Autor:innen und Institutionen

  • Department of Radiation Oncology, Yonsei Cancer Center, Yonsei University, College of Medicine & Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, Seoul, South Korea.
  • Korea Institute of Radiological and Medical Sciences, Seoul, South Korea.

ORCID-Profile: Dohyeon Yoo, Changhwan Kim, Min Cheol Han

Zitieren

APA

Yoo, D., Lee, H., Lee, S., Han, S., Kim, T., Kim, C., Han, M. C., & Kim, J. S. (2025). Investigation of an EPID‐based dose verification system for patient‐specific quality assurance in a harmony device. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70201

Vancouver

Yoo D, Lee H, Lee S, Han S, Kim T, Kim C, et al. Investigation of an EPID‐based dose verification system for patient‐specific quality assurance in a harmony device. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70201

BibTeX
@article{Yoo2025,
  author = {Yoo, Dohyeon and Lee, Hyunggun and Lee, Sangmin and Han, Soorim and Kim, Taeho and Kim, Changhwan and Han, Min Cheol and Kim, Jin Sung},
  title = {Investigation of an EPID‐based dose verification system for patient‐specific quality assurance in a harmony device},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70201}
}
Kostenlos starten →