An investigation into the feasibility and efficacy of stereotactic radiosurgery for 1–3 cm single brain lesions on the ring‐mounted Halcyon LINAC

Kate Hazelwood, Shane McCarthy iD, Josh Misa iD, David Castelvetere, William St Clair, Damodar Pokhrel iD

Journal of Applied Clinical Medical Physics · 2025-06 · 26 (6)

DOI: 10.1002/acm2.70105 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 09.04.2025. Verfasst von Kate Hazelwood und 5 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Graduate Program, Department of Radiation Medicine, University of Kentucky, Lexington, Kentucky, USA. Hauptthemen laut MEDLINE: Radiosurgery, Radiotherapy Planning, Computer-Assisted, Brain Neoplasms, Phantoms, Imaging. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

An evaluation of the accuracy, safety, and efficiency of the Halcyon ring delivery system (RDS) for stereotactic radiosurgery (SRS) treatment to relatively small (1–3 cm) brain lesions.

Methods

After completing the extensive in‐house quality assurance checks including Winston–Lutz test and independent dose verification via MD Anderson IROC SRS head phantom irradiation on Halcyon, fifteen brain SRS patients previously treated with a single dose of 20 Gy on TrueBeam (6MV‐FFF) with HyperArc geometry were retrospectively replanned on Halcyon (6MV‐FFF). Plan quality metrics including conformity index (CI), gradient index (GI), gradient distance (GD), PTV coverage, gross tumor volume (GTV) dose, heterogeneity index (HI), and doses to organs‐at‐risk (OAR) including normal brain dose were evaluated. Patient‐specific quality assurance (PSQA) and independent dose verification via in‐house Monte Carlo (MC) 2nd checks were performed.

Results

The Halcyon was able to provide highly conformal brain SRS plans. When compared to TrueBeam, CI, planning target volume (PTV) coverage, GTV dose (mean and minimum), HI, and doses to brainstem, optic pathway, and cochlea were statistically insignificant. Statistically significant increases in GI (3.76 vs. 3.25, p < 0.001), GD (0.56 cm vs. 0.48 cm, p = 0.001), and V 12Gy (5.5 cc vs. 4.6 cc, p = 0.014), on average using Halcyon versus TrueBeam was found, albeit clinically acceptable values for the majority of brain SRS cases. Halcyon plans provided statistically insignificant maximum dose to most adjacent OARs, though there was a statistically significant decrease in the maximum dose to the spinal cord (0.1 Gy vs. 0.4 Gy, p = 0.009). Halcyon beam‐on time increases by a factor of ∼2 ( p < 0.001). However, the faster patient setup on Halcyon results in a comparable estimated overall treatment time for both platforms. Plan deliverability and accuracy was ensured with PSQA (> 95% pass rate for 2%/2 mm clinical gamma criteria) results and MC 2nd check agreement within ± 5.0%.

Conclusions

Halcyon brain SRS plans provided a similar plan quality compared to HyperArc plans, although it demonstrated an inferior intermediate dose fall off thus slightly higher V 12Gy . This study suggests that Halcyon provides acceptable treatment for solitary relatively small brain lesions of 1–3 cm in diameter. Treatment of select patients on Halcyon will be started at our clinic and it is recommended that other clinics complete an end‐to‐end test, validate, and implement Halcyon SRS treatments at their practices, especially community cancer centers to provide high‐quality service to an underserved patient cohort.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryRadiotherapy Planning, Computer-AssistedBrain NeoplasmsPhantoms, ImagingOrgans at RiskParticle AcceleratorsHumansRadiotherapy DosageFeasibility StudiesRetrospective StudiesRadiotherapy, Intensity-ModulatedQuality Assurance, Health CareMonte Carlo Method

Stichwörter der Autor:innen

Halcyon LINACHyperArcSRSVMATco/non‐planar geometrysmall brain lesion

Autor:innen und Institutionen

  • Medical Physics Graduate Program, Department of Radiation Medicine, University of Kentucky, Lexington, Kentucky, USA.
  • Rhodes College, Physics, Biochemistry and Molecular Biology, Memphis, Tennessee, USA.

ORCID-Profile: Shane McCarthy, Josh Misa, Damodar Pokhrel

Zitieren

APA

Hazelwood, K., McCarthy, S., Misa, J., Castelvetere, D., St Clair, W., & Pokhrel, D. (2025). An investigation into the feasibility and efficacy of stereotactic radiosurgery for 1–3 cm single brain lesions on the ring‐mounted Halcyon LINAC. Journal of Applied Clinical Medical Physics, 26(6). https://doi.org/10.1002/acm2.70105

Vancouver

Hazelwood K, McCarthy S, Misa J, Castelvetere D, St Clair W, Pokhrel D. An investigation into the feasibility and efficacy of stereotactic radiosurgery for 1–3 cm single brain lesions on the ring‐mounted Halcyon LINAC. Journal of Applied Clinical Medical Physics. 2025;26(6). doi:10.1002/acm2.70105

BibTeX
@article{Hazelwood2025,
  author = {Hazelwood, Kate and McCarthy, Shane and Misa, Josh and Castelvetere, David and St Clair, William and Pokhrel, Damodar},
  title = {An investigation into the feasibility and efficacy of stereotactic radiosurgery for 1–3 cm single brain lesions on the ring‐mounted Halcyon LINAC},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  volume = {26},
  number = {6},
  doi = {10.1002/acm2.70105}
}
Kostenlos starten →