EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Clinical commissioning of a novel compact multi‐room pencil beam scanning proton therapy system

Eunsin Lee iD, Austin M Faught, Hyeri A Lee, Estelle Batin, Ahmet Ayan iD

Journal of Applied Clinical Medical Physics · 2026-03 · 27 (3)

DOI: 10.1002/acm2.70538 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 15.03.2026. Verfasst von Eunsin Lee und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, The Ohio State University, Columbus, Ohio, USA. Hauptthemen laut MEDLINE: Proton Therapy, Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging, Neoplasms. Die Arbeit nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To present the clinical commissioning of the world's first multi‐room Varian ProBeam 360° proton pencil beam scanning system.

Methods

The state‐of‐the‐art system includes two clinical treatment rooms with 360° rotating gantries, a superconducting cyclotron, an energy selection system, a beam transport system and scanning nozzles. These components deliver proton spots ranging from 3.97 to 30.03 g/cm 2 to arbitrarily shaped target volumes over a scanning area of 25 cm × 25 cm at isocenter. Proton beam ranges (R 80 ) were measured and verified independently in both gantry rooms to ensure agreement with beam specifications. Dosimetric parameters, including depth dose curves, in‐air spot profiles and dose per monitor unit (MU) as a function of energy, were measured and used to create a dose calculation model in the RayStation treatment planning system (TPS). Treatment plans with various sizes of spread‐out Bragg peaks (SOBPs) and simulated patient plans for a range of clinical sites were created and measured at various depths to validate the TPS's beam model accuracy. Additionally, beam matching between the two rooms was performed and validated.

Results

Across the full energy range of 69–218 MeV, the measured R 80 in both rooms were identical within 0.4 ± 0.2 mm deviation from the expected nominal range. In‐air spot sizes agreed within 5.7 ± 1.8% while outputs matched within 1.0 ± 0.5%. The average point dose difference was 0.2 ± 0.7% for over 80 measurements of SOBP validation plans from the TPS calculations and all planar dose measurements matched TPS calculations with over 90% of data passing a 2mm/2% gamma criterion. All patient plans were validated at a 3mm/3% criterion and demonstrated over 90% of data passing.

Conclusion

Both gantry systems were successfully commissioned for clinical use. Accurate measurements with robust validation ensured essential parameters for beam delivery and dosimetry were characterized for safe patient treatments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyRadiotherapy Planning, Computer-AssistedPhantoms, ImagingNeoplasmsParticle AcceleratorsRadiotherapy DosageHumansRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

PBS commissioningProBeam 360°Proton

Autor:innen und Institutionen

  • Department of Radiation Oncology, The Ohio State University, Columbus, Ohio, USA.

ORCID-Profile: Eunsin Lee, Ahmet Ayan

Zitieren

APA

Lee, E., Faught, A. M., Lee, H. A., Batin, E., & Ayan, A. (2026). Clinical commissioning of a novel compact multi‐room pencil beam scanning proton therapy system. Journal of Applied Clinical Medical Physics, 27(3). https://doi.org/10.1002/acm2.70538

Vancouver

Lee E, Faught AM, Lee HA, Batin E, Ayan A. Clinical commissioning of a novel compact multi‐room pencil beam scanning proton therapy system. Journal of Applied Clinical Medical Physics. 2026;27(3). doi:10.1002/acm2.70538

BibTeX
@article{Lee2026,
  author = {Lee, Eunsin and Faught, Austin M and Lee, Hyeri A and Batin, Estelle and Ayan, Ahmet},
  title = {Clinical commissioning of a novel compact multi‐room pencil beam scanning proton therapy system},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {3},
  doi = {10.1002/acm2.70538}
}
Kostenlos starten →