End‐to‐end film‐based characterization of gated delivery latencies in a spot‐scanned synchrotron‐based proton therapy system

Z S Stottler, J J Kruse, A J Deisher

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70826 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Z S Stottler und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA. Hauptthemen laut MEDLINE: Synchrotrons, Proton Therapy, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Temporal mismatches between gating signals and beam activation—beam‐on and beam‐off latencies—in gated particle beam therapy can compromise dose integrity. Yet standardized measurement techniques are lacking, and system‐specific characterization, particularly for synchrotron‐based proton systems, remains limited.

Purpose

To implement a novel strategy for precise end‐to‐end gating latency measurements of a spot‐scanned proton beam using Gafchromic film.

Methods

Gating latencies have been measured for a Hitachi ProBeat accelerator equipped with a Varian RPM gating system. EBT4 films were positioned within a stack of acrylic blocks on a QUASAR respiratory motion platform operated with a 1.5 cm amplitude and 12 bpm sinusoidal pattern. A stainless‐steel ball bearing (BB), positioned atop the film, facilitated localization. During irradiation, the BB produced a dose shadow on the film that indicated the phantom position. A proton plan consisting of a single 85.3 MeV spot was delivered to isocenter during a single gate‐on/off interval initiated near the 25% phase, to maximize phantom velocity. Fifty‐one measurements were made with a variety of gating intervals. Beam‐on latency and delivery time produced a separation between the BB‐shadow and the centroid of the Gaussian proton spot. Spot duration information was extracted from accelerator log files. BB‐shadow/spot separation is modeled analytically based on phantom motion enabling the direct calculation of the total beam‐on and beam‐off latencies. Accelerator latency was measured directly with an oscilloscope, enabling isolation of the relative contributions of the RPM gating system.

Results

An average beam‐on latency of ms is observed for the RPM+ProBeat system for the first gate of a delivery, which includes a known 69 ms beam extraction judgment time for the first proton spot in a spill from the Hitachi synchrotron. The corresponding average beam‐off latency is . The average gate‐on and gate‐off latencies of the RPM system are calculated as and , respectively. The associated average contributions of the accelerator to the beam‐on latencies are measured as (first gate in new spill) and (subsequent gates in existing spill) and similarly for beam‐off latency.

Conclusion

A novel, high‐precision method for gating‐latency measurements has been implemented using film to characterize gating latencies in a spot‐scanned particle beam. Commissioning and regular quality assurance methods to precisely characterize the temporal accuracy of a gating system ensure treatment accuracy and will become increasingly vital to consider for higher‐frequency applications, such as cardiac gating or FLASH‐capable systems.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SynchrotronsProton TherapyPhantoms, ImagingRadiotherapy Planning, Computer-AssistedFilm DosimetryParticle AcceleratorsHumansRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

RPMgatinglatencyprobeatproton

Autor:innen und Institutionen

  • Department of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA.

Zitieren

APA

Stottler, Z. S., Kruse, J. J., & Deisher, A. J. (2026). End‐to‐end film‐based characterization of gated delivery latencies in a spot‐scanned synchrotron‐based proton therapy system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70826

Vancouver

Stottler ZS, Kruse JJ, Deisher AJ. End‐to‐end film‐based characterization of gated delivery latencies in a spot‐scanned synchrotron‐based proton therapy system. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70826

BibTeX
@article{Stottler2026,
  author = {Stottler, Z S and Kruse, J J and Deisher, A J},
  title = {End‐to‐end film‐based characterization of gated delivery latencies in a spot‐scanned synchrotron‐based proton therapy system},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70826}
}
Kostenlos starten →