Repurposing DailyQA3 for an efficient and spot position sensitive daily quality assurance tool for proton therapy

Hong Qi Tan iD, Calvin Wei Yang Koh, Kah Seng Lew, Ping Lin Yeap, Clifford Ghee Ann Chua, James Kuan Huei Lee, Yun Ming Wong, Andrew Wibawa, Zubin Master, James Cheow Lei Lee, Sung Yong Park

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14348 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Hong Qi Tan und 10 weiteren Autor:innen. Beteiligte Einrichtung: Division of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore. Hauptthemen laut MEDLINE: Proton Therapy, Quality Assurance, Health Care, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Daily quality assurance is an integral part of a radiotherapy workflow to ensure the dose is delivered safely and accurately to the patient. It is performed before the first treatment of the day and needs to be time and cost efficient for a multiple gantries proton center. In this study, we introduced an efficient method to perform QA for output constancy, range verification, spot positioning accuracy and imaging and proton beam isocenter coincidence with DailyQA3.

Methods

A stepped acrylic block of specific dimensions is fabricated and placed on top of the DailyQA3 device. Treatment plans comprising of two different spread‐out Bragg peaks and five individual spots of 1.0 MU each are designed to be delivered to the device. A mathematical framework to measure the 2D distance between the detectors and individual spot is introduced and play an important role in realizing the spot positioning and centering QA. Lastly, a 5 months trends of the QA for two gantries are presented.

Results

The outputs are monitored by two ion chambers in the DailyQA3 and a tolerance of are used. The range of the SOBPs are monitored by the ratio of ion chamber signals and a tolerance of is used. Four diodes at from the central ion chambers are used for spot positioning QA, while the central ion chamber is used for imaging and proton beam isocenter coincidence QA. Using the framework, we determined the absolute signal threshold corresponding to the offset tolerance between the individual proton spot and the detector. A tolerances are used for both the positioning and centering QA. No violation of the tolerances is observed in the 5 months trends for both gantries.

Conclusion

With the proposed approach, we can perform four QA items in the TG224 within 10 min.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyQuality Assurance, Health CareRadiotherapy DosageRadiotherapy Planning, Computer-AssistedHumansNeoplasmsRadiotherapy, Intensity-ModulatedPhantoms, ImagingAlgorithmsRadiometry

Stichwörter der Autor:innen

DailyQA3pencil beam scanningproton therapyquality assurance

Autor:innen und Institutionen

  • Division of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
  • Oncology Academic Clinical Programme, Duke-NUS Medical School, Singapore, Singapore.
  • Division of Physics and Applied Physics, Nanyang Technological University, Singapore, Singapore.
  • Department of Physics, National University Singapore, Singapore, Singapore.

ORCID-Profile: Hong Qi Tan

Zitieren

APA

Tan, H. Q., Koh, C. W. Y., Lew, K. S., Yeap, P. L., Chua, C. G. A., Lee, J. K. H., Wong, Y. M., Wibawa, A., Master, Z., Lee, J. C. L., & Park, S. Y. (2024). Repurposing DailyQA3 for an efficient and spot position sensitive daily quality assurance tool for proton therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14348

Vancouver

Tan HQ, Koh CWY, Lew KS, Yeap PL, Chua CGA, Lee JKH, et al. Repurposing DailyQA3 for an efficient and spot position sensitive daily quality assurance tool for proton therapy. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14348

BibTeX
@article{Tan2024,
  author = {Tan, Hong Qi and Koh, Calvin Wei Yang and Lew, Kah Seng and Yeap, Ping Lin and Chua, Clifford Ghee Ann and Lee, James Kuan Huei and Wong, Yun Ming and Wibawa, Andrew and Master, Zubin and Lee, James Cheow Lei and Park, Sung Yong},
  title = {Repurposing DailyQA3 for an efficient and spot position sensitive daily quality assurance tool for proton therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14348}
}
Kostenlos starten →