Feasibility of proton daily QA prototype for pencil beam scanning carbon ion beam therapy
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70502 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Matthias Witt und 5 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Medical Physics and Radiation Protection, University of Applied Science, Gießen, Germany. Hauptthemen laut MEDLINE: Proton Therapy, Heavy Ion Radiotherapy, Radiotherapy Planning, Computer-Assisted, Quality Assurance, Health Care. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The complexity of beam delivery and the limited availability of dedicated quality assurance (QA) devices present unique challenges for carbon ion beam therapy. Most existing systems are designed for single‐parameter verification and have been adapted from photon or proton workflows, resulting in time‐consuming QA procedures. This study aimed to evaluate the feasibility of using a proton QA prototype device (DQA‐P) in a pencil beam scanning (PBS) carbon ion facility, with the aim of enabling fast, efficient, multiparameter QA.
Methods
The DQA‐P integrates 25 ionization chambers at three water‐equivalent depths, enabling the simultaneous measurement of multiple beam parameters, such as dose output, range, spot size, and spot position, in accordance with the recommendations of the AAPM TG‐224. Over 60 measurement sessions were performed, including 47 sessions involving intentional deviations in individual beam parameters, in order to assess sensitivity. The measurements were analyzed retrospectively using custom MATLAB scripts to evaluate accuracy, reproducibility, and the impact of setup uncertainties.
Results
The DQA‐P prototype successfully measured all QA parameters within clinically acceptable tolerances. Dose output measurements showed a standard deviation of 0.5% around the mean, with minimal systematic deviation. Spot position measurements exhibited a mean deviation of 0.16 mm, with a standard deviation of 0.32 mm. The device's sensitivity was demonstrated by its reliable detection of intentional positional offsets of 3 mm and spot size changes of 25%. Range validation using the proximal rise of the Bragg peak produced consistent results, albeit with lower accuracy than that achieved with standard multilayer ionization chambers. However, limitations were identified in spot size estimation due to chamber segmentation, and in range verification, as the device only captures three fixed depths.
Conclusions
The DQA‐P prototype demonstrated feasibility for daily QA in PBS carbon ion beam therapy, offering a short overall measurement time of approximately 5 min, efficient data acquisition, and simultaneous multiparameter verification within a single setup. This is a significant improvement on the conventional daily QA procedure, which uses two different detector types for independent dose and spot position measurements. In the standard workflow, both detectors must be positioned individually using the in‐room laser system, connected to the readout electronics and irradiated using the daily beam plan. Altogether, this procedure takes around 15–20 min. Although certain limitations remain—particularly in spot size accuracy and range determination—the device showed sufficient sensitivity and stability for daily QA purposes. Its use has the potential to streamline QA procedures, reduce beam time, and enhance workflow efficiency in carbon ion therapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Medical Physics and Radiation Protection, University of Applied Science, Gießen, Germany.
- Department of Radiotherapy and Radiation Oncology, Marburg University Hospital, Marburg, Germany.
- Marburg Ion-Beam Therapy Center (MIT), Marburg, Germany.
- SunNuclear, a Mirion Medical Company, Melbourne, Florida, USA.
- Biophysics Division, GSI Helmholtzzentrum für Schwerionenforschung, Darmstadt, Germany.
ORCID-Profile: Matthias Witt
Zitieren
APA
Witt, M., Flatten, V., Schönfeld, A., Baumann, K. S., Weber, U., & Zink, K. (2026). Feasibility of proton daily QA prototype for pencil beam scanning carbon ion beam therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70502
Vancouver
Witt M, Flatten V, Schönfeld A, Baumann KS, Weber U, Zink K. Feasibility of proton daily QA prototype for pencil beam scanning carbon ion beam therapy. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70502
BibTeX
@article{Witt2026,
author = {Witt, Matthias and Flatten, Veronika and Schönfeld, Andreas and Baumann, Kilian-Simon and Weber, Uli and Zink, Klemens},
title = {Feasibility of proton daily QA prototype for pencil beam scanning carbon ion beam therapy},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70502}
}
EUVIMED