A comprehensive pre‐clinical treatment quality assurance program using unique spot patterns for proton pencil beam scanning FLASH radiotherapy
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14400 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Pingfang Tsai und 8 weiteren Autor:innen. Beteiligte Einrichtung: New York Proton Center, New York, New York, USA. Hauptthemen laut MEDLINE: Quality Assurance, Health Care, Proton Therapy, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.
Abstract (Originalsprache)
Background
Quality assurance (QA) for ultra‐high dose rate (UHDR) irradiation is a crucial aspect in the emerging field of FLASH radiotherapy (FLASH‐RT). This innovative treatment approach delivers radiation at UHDR, demanding careful adoption of QA protocols and procedures. A comprehensive understanding of beam properties and dosimetry consistency is vital to ensure the safe and effective delivery of FLASH‐RT.
Purpose
To develop a comprehensive pre‐treatment QA program for cyclotron‐based proton pencil beam scanning (PBS) FLASH‐RT. Establish appropriate tolerances for QA items based on this study's outcomes and TG‐224 recommendations.
Methods
A 250 MeV proton spot pattern was designed and implemented using UHDR with a 215nA nozzle beam current. The QA pattern that covers a central uniform field area, various spot spacings, spot delivery modes and scanning directions, and enabling the assessment of absolute, relative and temporal dosimetry QA parameters. A strip ionization chamber array (SICA) and an Advanced Markus chamber were utilized in conjunction with a 2 cm polyethylene slab and a range ( R 80 ) verification wedge. The data have been monitored for over 3 months.
Results
The relative dosimetries were compliant with TG‐224. The variations of temporal dosimetry for scanning speed, spot dwell time, and spot transition time were within ± 1 mm/ms, ± 0.2 ms, and ± 0.2 ms, respectively. While the beam‐to‐beam absolute output on the same day reached up to 2.14%, the day‐to‐day variation was as high as 9.69%. High correlation between the absolute dose and dose rate fluctuations were identified. The dose rate of the central 5 × 5 cm 2 field exhibited variations within 5% of the baseline value (155 Gy/s) during an experimental session.
Conclusions
A comprehensive QA program for FLASH‐RT was developed and effectively assesses the performance of a UHDR delivery system. Establishing tolerances to unify standards and offering direction for future advancements in the evolving FLASH‐RT field.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- New York Proton Center, New York, New York, USA.
- Department of Radiation Oncology, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
- Department of Radiation Oncology, Montefiore Medical Center and Albert Einstein College of Medicine, Bronx, New York, USA.
ORCID-Profile: Pingfang Tsai, Chin-Cheng Chen, Wolfgang A Tomé, Haibo Lin
Förderung
- NCI NIH HHS
Zitieren
APA
Tsai, P., Yang, Y., Wu, M., Chen, C. C., Yu, F., Simone, C. B., Choi, J. I., Tomé, W. A., & Lin, H. (2024). A comprehensive pre‐clinical treatment quality assurance program using unique spot patterns for proton pencil beam scanning FLASH radiotherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14400
Vancouver
Tsai P, Yang Y, Wu M, Chen CC, Yu F, Simone CB, et al. A comprehensive pre‐clinical treatment quality assurance program using unique spot patterns for proton pencil beam scanning FLASH radiotherapy. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14400
BibTeX
@article{Tsai2024,
author = {Tsai, Pingfang and Yang, Yunjie and Wu, Mengjou and Chen, Chin-Cheng and Yu, Francis and Simone, Charles B and Choi, Jehee Isabelle and Tomé, Wolfgang A and Lin, Haibo},
title = {A comprehensive pre‐clinical treatment quality assurance program using unique spot patterns for proton pencil beam scanning FLASH radiotherapy},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14400}
}
EUVIMED