Evaluating equivalent square field size definitions for rectangular small fields

Ignasi Méndez iD, Mateb Al Khalifa iD, Haya Aljuaid iD, Božidar Casar

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70500 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Ignasi Méndez und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Dosimetry and Quality of Radiological Procedures, Institute of Oncology Ljubljana, Ljubljana, Slovenia. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging, Algorithms, Film Dosimetry. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Slovenian Research and Innovation Agency.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In the IAEA TRS‐483 Code of Practice (CoP), rectangular small field sizes are approximated to equivalent square small fields using the definition proposed by Cranmer‐Sargison et al. However, the CoP estimates the uncertainties associated with this formula only for rectangular fields with dimensions within the range .

Purpose

The objective of the present study was to compare the accuracy of the Cranmer‐Sargison definition with other formulas for equivalent square small fields in the context of measuring field output factors (FOFs) for rectangular small fields, both within and outside the range covered by the CoP.

Methods

Measurements were conducted using Gafchromic EBT4 radiochromic films. The models compared included Cranmer‐Sargison, Sterling, Superellipse, Sterling‐Partial Superellipse, Sterling‐Superellipse, Vadash and Bjärngard, and Fogliata. The most accurate definition of equivalent square field size was identified as the one yielding the lowest discrepancy between measured and analytical values, with the log‐likelihood of the measurements selected as the metric. Analytical values were derived using the function introduced by Sauer and Wilbert, which relates FOFs to equivalent square field sizes.

Results

The best results were achieved with the Fogliata model, followed in terms of accuracy by the Sterling‐Partial Superellipse model. The Sterling‐Superellipse and Vadash and Bjärngard models came next. It should be noted that the Sterling‐Partial Superellipse and Sterling‐Superellipse models rely solely on the geometric shape of the irradiation field size, whereas the Fogliata and Vadash and Bjärngard models incorporate a fitting parameter. The Sterling definition, while less accurate than these models, improved upon the Cranmer‐Sargison definition and retained computational simplicity. Finally, the Cranmer‐Sargison and Superellipse models exhibited the largest discrepancies.

Conclusions

This study identified several definitions of equivalent square small field size that could refine the IAEA TRS‐483 CoP by improving the accuracy of field output correction factors for rectangular small fields. Among these definitions, the Fogliata model obtained the best results.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedPhantoms, ImagingAlgorithmsFilm DosimetryModels, TheoreticalRadiotherapy DosageHumans

Stichwörter der Autor:innen

equivalent square field sizefield output factorsradiochromic filmssmall fields dosimetry

Autor:innen und Institutionen

  • Department of Dosimetry and Quality of Radiological Procedures, Institute of Oncology Ljubljana, Ljubljana, Slovenia.
  • Virginia Commonwealth University, Richmond, Virginia, USA.
  • Department of Radiation Medicine, Northwell Health, Lake Success, New York, USA.

ORCID-Profile: Ignasi Méndez, Mateb Al Khalifa, Haya Aljuaid

Förderung

  • Slovenian Research and Innovation Agency

Zitieren

APA

Méndez, I., Al Khalifa, M., Aljuaid, H., & Casar, B. (2026). Evaluating equivalent square field size definitions for rectangular small fields. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70500

Vancouver

Méndez I, Al Khalifa M, Aljuaid H, Casar B. Evaluating equivalent square field size definitions for rectangular small fields. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70500

BibTeX
@article{Mndez2026,
  author = {Méndez, Ignasi and Al Khalifa, Mateb and Aljuaid, Haya and Casar, Božidar},
  title = {Evaluating equivalent square field size definitions for rectangular small fields},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70500}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche