Lateral response artifact correction method using image stitching technique in radiochromic film dosimetry

Hideharu Miura, Masanori Miyazawa, Shuichi Ozawa, Tsubasa Enosaki, Masayuki Kagemoto

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14373 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Hideharu Miura und 4 weiteren Autor:innen. Beteiligte Einrichtung: Hiroshima High-Precision Radiotherapy Cancer Center, Hiroshima, Japan. Hauptthemen laut MEDLINE: Film Dosimetry, Artifacts, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: JSPS KAKENHI.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Lateral response artifact (LRA) is caused by the interaction between film and flatbed scanner in the direction perpendicular to the scanning direction. This can significantly affect the accuracy of patient‐specific quality assurance (QA) in cases involving large irradiation fields. We hypothesized that by utilizing the central area of the flatbed scanner, where the magnitude of LRA is relatively small, the LRA could be mitigated effectively. This study proposes a practical solution using the image‐stitching technique to correct LRA for patient‐specific QA involving large irradiation fields.

Methods

Gafchromic™ EBT4 film and Epson Expression ES‐G11000 flatbed scanner were used in this study. The image‐stitching algorithm requires a spot between adjacent images to combine them. The film was scanned at three locations on a flatbed scanner, and these images were combined using the image‐stitching technique. The combined film dose was then calculated and compared with the treatment planning system (TPS)‐calculated dose using gamma analysis (3%/2 mm). Our proposed LRA correction was applied to several films exposed to 18 × 18 cm 2 open fields at doses of 200, 400, and 600 cGy, as well as to four clinical Volumetric Modulated Arc Therapy (VMAT) treatment plans involving large fields.

Results

For doses of 200, 400, and 600 cGy, the gamma analysis values with and without LRA corrections were 95.7% versus 67.8%, 95.5% versus 66.2%, and 91.8% versus 35.9%, respectively. For the clinical VMAT treatment plan, the average pass rate ± standard deviation in gamma analysis was 94.1% ± 0.4% with LRA corrections and 72.5% ± 1.5% without LRA corrections.

Conclusions

The effectiveness of our proposed LRA correction using the image‐stitching technique was demonstrated to significantly improve the accuracy of patient‐specific QA for VMAT treatment plans involving large irradiation fields.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Film DosimetryArtifactsRadiotherapy Planning, Computer-AssistedRadiotherapy DosageAlgorithmsImage Processing, Computer-AssistedQuality Assurance, Health CareRadiotherapy, Intensity-ModulatedPhantoms, ImagingHumans

Stichwörter der Autor:innen

film dosimetryimage‐stitching techniquelateral response artifact

Autor:innen und Institutionen

  • Hiroshima High-Precision Radiotherapy Cancer Center, Hiroshima, Japan.
  • Department of Radiation Oncology, Institute of Biomedical & Health Sciences, Hiroshima University, Hiroshima, Japan.
  • R-TECH.INC, Tokyo, Japan.

Förderung

  • JSPS KAKENHI

Zitieren

APA

Miura, H., Miyazawa, M., Ozawa, S., Enosaki, T., & Kagemoto, M. (2024). Lateral response artifact correction method using image stitching technique in radiochromic film dosimetry. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14373

Vancouver

Miura H, Miyazawa M, Ozawa S, Enosaki T, Kagemoto M. Lateral response artifact correction method using image stitching technique in radiochromic film dosimetry. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14373

BibTeX
@article{Miura2024,
  author = {Miura, Hideharu and Miyazawa, Masanori and Ozawa, Shuichi and Enosaki, Tsubasa and Kagemoto, Masayuki},
  title = {Lateral response artifact correction method using image stitching technique in radiochromic film dosimetry},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14373}
}
Kostenlos starten →