Clinical use of Gafchromic EBT4 film for in vivo dosimetry for total body irradiation
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.14574 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Emily Draeger und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Therapeutic Radiology, Yale University School of Medicine, New Haven, Connecticut, USA. Hauptthemen laut MEDLINE: Film Dosimetry, Whole-Body Irradiation, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.
Abstract (Originalsprache)
Purpose
In vivo dosimetry is a common requirement to validate dose accuracy/uniformity in total body irradiation (TBI). Several detectors can be used for in vivo dosimetry, including thermoluminescent dosimeters (TLDs), diodes, ion chambers, optically stimulated luminescent dosimeters (OSLDs), and film. TLDs are well established for use in vivo but required expertise and clinical system availability may make them impractical for multifractionated TBI. OSLDs offer quick readout, but recalls have restricted their use. The purpose of this work was to validate the newly available Gafchromic EBT4 film for TBI in vivo dosimetry.
Methods
Film calibration curves were created under standard conditions (6MV/15MV, 1.5/3.0 cm depth, 100 cm source‐to‐surface distance (SSD), 10 × 10 cm 2 field), and films were scanned at several time points (0.5–24 h) to determine the shortest development time that yielded accurate dose measurements. 4 × 4 cm 2 films were placed under 1.5 cm thick bolus on the anterior and posterior sides of a solid water phantom to measure entrance and exit dose under TBI conditions (∼600 cm SSD, 39.5 × 39.5 cm 2 field, 6 MV/15 MV). These measurements were compared to ion chamber and diode readings for validation. Film measurements were also compared to OSLD measurements for three TBI patients.
Results
The shortest development time that resulted in accurate dosimetry and allowed for adequate physician review time was 4 h (± 4% dose accuracy). Film entrance and exit dose measurements were within ± 3.8% of ion chamber and diode readings for 6MV and 15MV beams. Patient film measurements were within ∼ ± 5% for the majority of anatomical measurement locations; however, film and OSLD readings for some anatomic locations deviated by > 10%.
Conclusions
These results indicate that EBT4 film can be utilized for accurate in vivo dosimetry for TBI patients and shows good agreement with diode and ion chamber measurements. Further investigation into film and OSLD differences was not possible due to OSLD recalls.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Therapeutic Radiology, Yale University School of Medicine, New Haven, Connecticut, USA.
- Department of Radiation Physics, MD Anderson Cancer Center, Houston, Texas, USA.
- Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
Förderung
- NCI NIH HHS
Zitieren
APA
Draeger, E., Guan, F., Lee, M. Y., Han, D. Y., Donahue, W., & Chen, Z. J. (2025). Clinical use of Gafchromic EBT4 film for in vivo dosimetry for total body irradiation. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14574
Vancouver
Draeger E, Guan F, Lee MY, Han DY, Donahue W, Chen ZJ. Clinical use of Gafchromic EBT4 film for in vivo dosimetry for total body irradiation. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.14574
BibTeX
@article{Draeger2025,
author = {Draeger, Emily and Guan, Fada and Lee, Min-Young and Han, Dae Yup and Donahue, William and Chen, Zhe Jay},
title = {Clinical use of Gafchromic EBT4 film for in vivo dosimetry for total body irradiation},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.14574}
}
EUVIMED