EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Recommendations for using analytical anisotropic algorithm and AcurosXB for epidermal dose calculations in breast radiotherapy from an in vivo Gafchromic film study

Aria Malhotra iD, Emilie E Carpentier, Cheryl Duzenli iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14416 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Aria Malhotra und 2 weiteren Autor:innen. Beteiligte Einrichtung: BC Cancer, Vancouver, British Columbia, Canada. Hauptthemen laut MEDLINE: Algorithms, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Breast Neoplasms. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and purpose This study recommends clinical epidermal dose calculation methods based on in‐vivo film measurements and registered skin dose distributions with the Eclipse (Varian Medical Systems) treatment planning system's Analytical Anisotropic Algorithm (AAA) and Acuros XB (AXB) dose calculation algorithms. Materials and methods Eighteen AAA V13.6 breast plans were recalculated using AXB (dose to medium) V13.5 with the same beam parameters and monitor units as in the original plans. These are compared against in‐vivo Gafchromic film measurements from the lateral and inferior breast regions. Three skin structures in the treatment planning system are evaluated: a surface layer of voxels of the body contour, a 0.2 cm internal skin rind, and a 0.5 cm internal skin rind.

Results

Systematic shifts are demonstrated between the film measurements of skin dose and the Eclipse dose calculations. On average, the dose to the surface layer of pixels is underestimated by AAA by 8% and overestimated by AXB by 3%. A 5 mm skin rind extended into the body can increase epidermal dose calculations on average by 8% for AAA and 4% for AXB.

Conclusion

This is the first study to register in‐vivo skin dose distributions in the breast to the treatment planning system for comparison. Based on the results from this study it is recommended that epidermal dose is calculated with a 0.5 cm skin rind for the AAA algorithm and with rind thickness up to 0.2 cm for the AXB algorithm.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsRadiotherapy Planning, Computer-AssistedRadiotherapy DosageBreast NeoplasmsFilm DosimetryRadiotherapy, Intensity-ModulatedHumansFemaleAnisotropyEpidermisOrgans at RiskPhantoms, Imaging

Stichwörter der Autor:innen

Eclipse AAA and AXB algorithmsGafChromic filmbreast radiotherapyin vivo skin dose

Autor:innen und Institutionen

  • BC Cancer, Vancouver, British Columbia, Canada.
  • Department of Physics and Astronomy, University of British Columbia, Vancouver, British Columbia, Canada.
  • Department of Surgery, Division of Radiation Oncology and Experimental Radiotherapeutics, University of British Columbia, Vancouver, British Columbia, Canada.

ORCID-Profile: Aria Malhotra, Cheryl Duzenli

Zitieren

APA

Malhotra, A., Carpentier, E. E., & Duzenli, C. (2024). Recommendations for using analytical anisotropic algorithm and AcurosXB for epidermal dose calculations in breast radiotherapy from an in vivo Gafchromic film study. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14416

Vancouver

Malhotra A, Carpentier EE, Duzenli C. Recommendations for using analytical anisotropic algorithm and AcurosXB for epidermal dose calculations in breast radiotherapy from an in vivo Gafchromic film study. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14416

BibTeX
@article{Malhotra2024,
  author = {Malhotra, Aria and Carpentier, Emilie E and Duzenli, Cheryl},
  title = {Recommendations for using analytical anisotropic algorithm and AcurosXB for epidermal dose calculations in breast radiotherapy from an in vivo Gafchromic film study},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14416}
}
Infokarte | EUVIMED