Performance evaluation and quantitative comparison of two 4DCT imaging respiratory systems using deformable image registration

Ali Al-Zein, Rawan H Naim, Wassim Jalbout, Bilal H Shahine iD

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70279 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Ali Al-Zein und 3 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Oncology Department, American University of Beirut Medical Center, Beirut, Lebanon. Hauptthemen laut MEDLINE: Four-Dimensional Computed Tomography, Phantoms, Imaging, Image Processing, Computer-Assisted, Respiration. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Improved accuracy in 4DCT imaging and precise targeting of tumors contribute to more effective and targeted radiation therapy. This study focuses on evaluating the accuracy of utilizing the GateCT (VisionRT Ltd, London, United Kingdom) in comparison with a pressure sensor system (ANZAI Medical Co., Ltd., Shinagawa, Tokyo) to provide 4DCT with respiratory information.

Methods

A dynamic breathing phantom enclosing three spheres (A, P, and R) of different densities was enrolled to produce breathing patterns tracked by the two systems. Image sets for three breathing phases obtained based on GateCT and ANZAI systems were analyzed using deformable registration by deforming the three‐phase image sets with the static image sets. Our deformable registration approach revealed how far different phase image sets were from the quantified by various metrics, such as dice similarity coefficient (DSC), mean surface distance (MSD), absolute volume estimation, mean Jacobian, and Warp.

Results

Results indicated DSC values greater than 0.90 across all phases and spheres for both respiratory systems, with mean DSC values for spheres A, P, and R of 0.980 versus 0.977, 0.977 versus 0.976, and 0.977 versus 0.976 for GateCT and ANZAI systems, respectively. MSDs for both systems were consistently less than 2 mm across all spheres and phases. Furthermore, the mean volume estimation error for both systems, relative to the static, exhibited statistical insignificance ( p > 0.05). Friedman test revealed significant differences in median Jacobian, and median Warping between the two systems ( p < 0.05).

Conclusions

In addition to the effectiveness of deformable image registration in the quantification of respiratory system performance, both systems exhibited comparable performance in providing 4DCT with respiratory information.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Four-Dimensional Computed TomographyPhantoms, ImagingImage Processing, Computer-AssistedRespirationRadiotherapy Planning, Computer-AssistedLung NeoplasmsRespiratory-Gated Imaging TechniquesHumansAlgorithmsRadiotherapy, Intensity-ModulatedRadiotherapy Dosage

Stichwörter der Autor:innen

4DCT image reconstructionGateCTdeformable image registrationpressure sensor ANZAI

Autor:innen und Institutionen

  • Radiation Oncology Department, American University of Beirut Medical Center, Beirut, Lebanon.

ORCID-Profile: Bilal H Shahine

Zitieren

APA

Al-Zein, A., Naim, R. H., Jalbout, W., & Shahine, B. H. (2025). Performance evaluation and quantitative comparison of two 4DCT imaging respiratory systems using deformable image registration. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70279

Vancouver

Al-Zein A, Naim RH, Jalbout W, Shahine BH. Performance evaluation and quantitative comparison of two 4DCT imaging respiratory systems using deformable image registration. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70279

BibTeX
@article{AlZein2025,
  author = {Al-Zein, Ali and Naim, Rawan H and Jalbout, Wassim and Shahine, Bilal H},
  title = {Performance evaluation and quantitative comparison of two 4DCT imaging respiratory systems using deformable image registration},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70279}
}
Kostenlos starten →