Localization accuracy of 6‐second CBCT for lung IGRT with various breathing patterns

Jihye Koo iD, Gage Redler iD, Vladimir Semenenko, Stephen A Rosenberg, Emily Keit, Jacqueline M Andreozzi

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70130 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Jihye Koo und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, H. Lee Moffitt Cancer Center and Research Institute, Tampa, Florida, USA. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Lung Neoplasms, Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The 6‐second cone beam computed tomography (CBCT) acquisition of the Ethos HyperSight (Varian Medical Systems, Inc. Palo Alto, CA, USA) on‐board imaging system offers benefits, but could be too fast to accurately capture an average target position in a free‐breathing lung cancer patient. This study aimed to ascertain whether a 6‐second acquisition is appropriate for regularly breathing patients with varying respiration periods. Additionally, breathing patterns that could lead to clinically impactful inaccuracies for image‐guided radiation therapy (IGRT) shifts were investigated.

Methods

Nine regular (sinusoidal) breathing patterns with different respiration rates (8–20 breaths‐per‐minute) and amplitudes, along with five irregular breathing patterns including a gasp/cough scenario, were tested using a respiratory motion phantom with a 3 cm diameter spherical target. Once the phantom was aligned using the external chassis to remove any default shifts, the 6‐second CBCTs with Acuros reconstruction were acquired. Rigid registrations were performed using the 4DCT simulation average/untagged reconstructed image as reference to align to target (simulating clinical lung IGRT setup). Sixty‐second CBCTs were also tested to verify the average offsets with conventional practice. The IGRT shifts were compared to 5 mm, a PTV margin typically used for lung tumors.

Results

Minimal ( 0.05) between the sizes of shifts made with the two CBCT durations.

Conclusion

The 6‐second CBCT can effectively and reliably localize a mobile target for regular and moderately‐irregular breathers. Cautions should be used for highly irregular breathers, regardless of the duration of CBCT acquisition.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyLung NeoplasmsRadiotherapy Planning, Computer-AssistedPhantoms, ImagingRespirationRadiotherapy, Image-GuidedImage Processing, Computer-AssistedHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedFour-Dimensional Computed Tomography

Stichwörter der Autor:innen

4DCTCBCTIGRTlocalizationrespiratory motion

Autor:innen und Institutionen

  • Department of Radiation Oncology, H. Lee Moffitt Cancer Center and Research Institute, Tampa, Florida, USA.

ORCID-Profile: Jihye Koo, Gage Redler

Zitieren

APA

Koo, J., Redler, G., Semenenko, V., Rosenberg, S. A., Keit, E., & Andreozzi, J. M. (2025). Localization accuracy of 6‐second CBCT for lung IGRT with various breathing patterns. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70130

Vancouver

Koo J, Redler G, Semenenko V, Rosenberg SA, Keit E, Andreozzi JM. Localization accuracy of 6‐second CBCT for lung IGRT with various breathing patterns. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70130

BibTeX
@article{Koo2025,
  author = {Koo, Jihye and Redler, Gage and Semenenko, Vladimir and Rosenberg, Stephen A and Keit, Emily and Andreozzi, Jacqueline M},
  title = {Localization accuracy of 6‐second CBCT for lung IGRT with various breathing patterns},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70130}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche