Log file‐based quality assurance method for respiratory gating system
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70101 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Wonjoong Cheon und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Quality Assurance, Health Care, Phantoms, Imaging, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Research Fund of Seoul St.Mary's Hospital, The Catholic University of Korea, Korea Electrotechnology Research Institute, National Research Council of Science and Technology.
Abstract (Originalsprache)
Background
As medical linear accelerator technology advances, enabling higher dose rate deliveries, hypofractionation regimens has increased. This necessitates respiratory gating systems that synchronize radiation delivery with tumor position, requiring simple rigorous quality assurance (QA) to ensure treatment accuracy and patient safety.
Purpose
This study aimed to propose log‐based QA for respiratory‐gated radiation therapy using the respiratory gating system and treatment machine.
Methods
4D CT scans were performed with a Varian motion phantom using a Varian Respiratory Gating for Scanner (RGSC). A treatment plan using 25%–75% respiratory phases with 100 MU was created and delivered to a solid water phantom. Treatment logs containing respiratory signals, beam on/off flags, and frame information were extracted from the treatment planning system's offline review. Log file analyses were conducted using in‐house softwares to assess temporal synchronization between respiratory phases and beam triggers. Output measurements using a calibrated ion chamber (FC65G) were performed to evaluate dosimetric accuracy. Additionally, EPID images were acquired in cine mode and analyzed frame‐by‐frame to independently verify beam delivery timing.
Results
Log file analysis revealed precise temporal synchronization, with mean time differences of 0.03 s ± 0.05 s between the planned 25% phase and beam‐on, and −0.04 s ± 0.05 s between 75% phase and beam‐off. The log‐derived beam‐on duration (2.61 s ± 0.02 s) closely matched the planned duration (2.66 s ± 0.00 s). Three‐month log data showed consistent temporal accuracy, with trigger‐on times remaining stable at 2.60 s ± 0.01 s across all measurements. Supporting ion chamber measurements confirmed dosimetric agreement between gating and non‐gating modes (difference: 0.05 cGy ± 0.09 cGy)
Conclusions
The proposed log file‐based QA method demonstrated high accuracy and reproducibility in assessing respiratory gating performance. This approach provides an efficient, objective method for standardizing QA procedures in respiratory‐gated radiation therapy, enhancing treatment accuracy and patient safety.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
- Department of Radiation Oncology, Seoul St. Mary's Hospital, The Catholic University of Korea, Seoul, Republic of Korea.
- Department of Biomedicine & Helath Sciences, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
- Division of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
- Oncology Academic Clinical Programme, Duke-NUS Medical School, Singapore, Singapore.
- School of Physics and Mathematical Science, Nanyang Technological University, Singapore, Singapore.
ORCID-Profile: Wonjoong Cheon
Förderung
- Research Fund of Seoul St.Mary's Hospital, The Catholic University of Korea
- Korea Electrotechnology Research Institute
- National Research Council of Science and Technology
Zitieren
APA
Cheon, W., Lee, Y. K., Seol, Y., Park, C. B., Tan, H. Q., Choi, K. H., Kang, Y. N., & Choi, B. O. (2025). Log file‐based quality assurance method for respiratory gating system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70101
Vancouver
Cheon W, Lee YK, Seol Y, Park CB, Tan HQ, Choi KH, et al. Log file‐based quality assurance method for respiratory gating system. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70101
BibTeX
@article{Cheon2025,
author = {Cheon, Wonjoong and Lee, Young Kyu and Seol, Yunji and Park, Chan-Beom and Tan, Hong Qi and Choi, Kyu Hye and Kang, Young-Nam and Choi, Byung Ock},
title = {Log file‐based quality assurance method for respiratory gating system},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70101}
}
EUVIMED