The edge visualization metric: Quantifying the improvement of lung SBRT target definition with 4D CBCT

Colton Baley, Sam Andersen, Clark Anderson, Shraddha Dalwadi, Daniel L Saenz iD

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70114 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Colton Baley und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Mays Cancer Center at UT Health San Antonio, San Antonio, Texas, USA. Hauptthemen laut MEDLINE: Lung Neoplasms, Four-Dimensional Computed Tomography, Phantoms, Imaging, Cone-Beam Computed Tomography. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Four‐dimensional cone‐beam CT (4D CBCT) incorporates oversampling of 3D data to reconstruct multi‐phase CBCT data sets representing distinct phases of the breathing cycle based on a diaphragmatic correlate of respiratory motion. Motion artifacts and blurring can be reduced relative to three‐dimensional cone‐beam (3D CBCT), allowing clinicians to better visualize motion of targets. To quantitatively understand the degree to which target visualization is improved by 4D CBCT, an edge visualization metric (EVM) has been developed to describe the change in voxel intensities at the edge of targets in 4D CBCT maximum intensity projection images relative to 3D CBCT images.

Methods

The EVM describes the median distance where voxel intensities drop from 80% to 20% of target voxel values. The EVM was evaluated in a phantom study with a CIRS dynamic thorax phantom and with eleven on‐treatment lung SBRT patients.

Results

In the phantom study, the EVM was improved for 4D CBCT relative to 3D CBCT for one‐cm targets (2.43 ± 0.22 mm vs. 2.67 ± 0.31 mm, p = 0.04) and for 2‐cm targets (2.60 ± 0.35 mm vs. 3.46 ± 1.03 mm, p = 0.02). In patients, the EVM was 3.59 ± 1.01 mm vs. 4.25 ± 1.24 mm ( p < 0.05).

Conclusions

When evaluating an imaging acquisition's degree of motion blurring and ability to delineate target edges, EVM may provide a less biased way to evaluate edge detection in the presence of motion when compared to traditional methods.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lung NeoplasmsFour-Dimensional Computed TomographyPhantoms, ImagingCone-Beam Computed TomographyRadiosurgeryRadiotherapy Planning, Computer-AssistedImage Processing, Computer-AssistedCarcinoma, Non-Small-Cell LungHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedArtifactsRespirationAlgorithmsMovement

Stichwörter der Autor:innen

CBCTSBRTlungradiation therapy

Autor:innen und Institutionen

  • Department of Radiation Oncology, Mays Cancer Center at UT Health San Antonio, San Antonio, Texas, USA.

ORCID-Profile: Daniel L Saenz

Zitieren

APA

Baley, C., Andersen, S., Anderson, C., Dalwadi, S., & Saenz, D. L. (2025). The edge visualization metric: Quantifying the improvement of lung SBRT target definition with 4D CBCT. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70114

Vancouver

Baley C, Andersen S, Anderson C, Dalwadi S, Saenz DL. The edge visualization metric: Quantifying the improvement of lung SBRT target definition with 4D CBCT. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70114

BibTeX
@article{Baley2025,
  author = {Baley, Colton and Andersen, Sam and Anderson, Clark and Dalwadi, Shraddha and Saenz, Daniel L},
  title = {The edge visualization metric: Quantifying the improvement of lung SBRT target definition with 4D CBCT},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70114}
}
Kostenlos starten →