Evaluating artifact‐free four‐dimensional computer tomography with 16 cm detector array

Inhwan Yeo iD, Wei Nie, Jiajin Fan, Mindy Joo, Michael Correa, Qianyi Xu

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70056 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Inhwan Yeo und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Advanced Radiation and Proton Therapy, INOVA Schar Cancer Institute, Fairfax, Virginia, USA. Hauptthemen laut MEDLINE: Four-Dimensional Computed Tomography, Phantoms, Imaging, Lung Neoplasms, Artifacts. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate a 16 cm‐array axial four‐dimensional computer tomography (4DCT) in comparison with a 4 cm‐array 4DCT in the presence of respiration irregularity.

Method

Ten traces of lung tumor motion from CyberKnife treatments were imported to move the lung cylinder, containing a spherical target, of a phantom. Images were acquired for the lung that moved to each of the 10‐positions/phases (1) step‐wisely by nominal helical scan at each movement (ground truth), (2) continuously by 4D scan with the 16 cm array, and (3) the same with the 4 cm array, involving table shift. Irregularities, consisting of baseline shift and/or amplitude change of the traces in their second periods, affected #3 scan only in its second table position. The full‐widths at half maximum of the target in the direction of the motion were determined on the average (Ave) CT, maximum‐intensity (Mip) CT, and a phase (MP) CT that is associated with the maximum error, comparing #2 and 3 with #1. Three tumor‐shaped targets were also imaged, and overlap ratios of them from #2 and 3 with the targets from #1 were inter‐compared. Hounsfield unit (HU)s of the targets were also compared.

Results

The average difference in the spherical‐target length between #2 and #1 was found to be 0.28 ± 0.15 cm on AveCT, 0.00 ± 0.18 cm on MipCT, and 0.07 ± 0.06 cm on MPCT, showing agreement. The average difference between #3 and #1 was 0.34 ± 0.23 cm on AveCT, 0.48 ± 0.31 cm on MipCT, and 0.56 ± 0.50 cm on MPCT, showing disagreement. The overlap ratios were better with #2 than with #3 for all tumor‐shaped targets in each phase CT and MipCT, but they were not perfect for #2 due to motion averaging and phase sorting limitations. The differences in HUs were smaller with #2 than with #3, but not fully satisfactory with #2.

Conclusion

4DCT with the 16 cm array needs to be used to minimize the impact of the irregularity.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Four-Dimensional Computed TomographyPhantoms, ImagingLung NeoplasmsArtifactsRadiotherapy Planning, Computer-AssistedImage Processing, Computer-AssistedRadiosurgeryHumansMovementRespirationRadiotherapy, Intensity-ModulatedRadiotherapy Dosage

Stichwörter der Autor:innen

16 cm long detector4DCTartifactsrespiration irregularity

Autor:innen und Institutionen

  • Department of Advanced Radiation and Proton Therapy, INOVA Schar Cancer Institute, Fairfax, Virginia, USA.
  • Radiation Oncology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.

ORCID-Profile: Inhwan Yeo

Zitieren

APA

Yeo, I., Nie, W., Fan, J., Joo, M., Correa, M., & Xu, Q. (2025). Evaluating artifact‐free four‐dimensional computer tomography with 16 cm detector array. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70056

Vancouver

Yeo I, Nie W, Fan J, Joo M, Correa M, Xu Q. Evaluating artifact‐free four‐dimensional computer tomography with 16 cm detector array. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70056

BibTeX
@article{Yeo2025,
  author = {Yeo, Inhwan and Nie, Wei and Fan, Jiajin and Joo, Mindy and Correa, Michael and Xu, Qianyi},
  title = {Evaluating artifact‐free four‐dimensional computer tomography with 16 cm detector array},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70056}
}
Kostenlos starten →