Impact of tumor motion on target delineation and dose calculation accuracy using rapid‐acquisition HyperSight CBCT in online adaptive radiotherapy
Journal of Applied Clinical Medical Physics · 2026-01 · 27 (1)
DOI: 10.1002/acm2.70428 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 28.12.2025. Verfasst von Yi-Fang Wang und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Vagelos College of Physicians Surgeons, Columbia University Irving Medical Center, New York, New York, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Cone-Beam Computed Tomography, Phantoms, Imaging, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
HyperSight CBCT offers rapid image acquisition and enhanced image quality, supporting its integration into online adaptive radiotherapy (OART). However, its performance under respiratory‐induced tumor motion requires further evaluation.
Purpose
This study evaluates the feasibility of using rapid‐acquisition HyperSight CBCT for internal target volume (ITV) delineation and dose calculation in the setting of respiratory motion, with implications for OART workflows.
Methods
A dynamic thoracic phantom simulating superior‐inferior tumor motion (5–25 mm amplitude from the motion center, 6‐second cycle time) was imaged using Varian's HyperSight CBCT on the Ethos platform. Four reconstruction algorithms were evaluated: filtered back projection (FDK), iterative CBCT (iCBCT), iCBCT with Acuros, and metal artifact reduction (MAR). Experiments included variation of cycle times (4, 6, and 8 s), breathing patterns (Cos 6 , sinusoidal, and hysteresis), and repeat acquisitions to assess the impact of scan initiation time relative to the respiratory cycle. In addition to fast scans, slow acquisitions (60‐second cycle) were performed for further evaluation. Reference datasets included 4DCT maximum intensity projection (MIP) and average intensity projection (AIP) for ITV delineation and HU comparison. Volumetric modulated arc therapy plans were created on AIP images and recalculated on HyperSight images. Dose distributions were compared using gamma analysis (1%/1 mm, 10% threshold).
Results
For the Cos 6 breathing pattern with a 6‐second cycle time, HyperSight CBCT produced ITV volumes and HU values comparable to 4DCT at small respiratory amplitudes (≤10 mm), with Dice similarity coefficients exceeding 0.8 and dose calculations passing 1%/1 mm gamma analysis. At larger amplitudes (≥15 mm), ITVs were underestimated to 28–40% of the reference values, with Dice coefficients falling below 0.45 and increasing image distortion. Longer cycle times (8s) and irregular breathing patterns (hysteresis) further reduced trajectory visualization, while more uniform motion (sinusoidal) improved trajectory coverage but still showed image deformation at 15mm amplitude. Repeat acquisitions demonstrated varying image representations at different scan initiation times, particularly with the longer 8‐second cycle. Slow‐scan CBCT achieved close agreement with 4DCT MIP and AIP in ITV volumes and HU values.
Conclusions
HyperSight CBCT is feasible for OART in patients with limited respiratory motion, regular breathing pattern and short breathing cycle time. For larger excursions, slow scans or motion management strategies may be required to ensure accurate target delineation and dose calculation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Vagelos College of Physicians Surgeons, Columbia University Irving Medical Center, New York, New York, USA.
ORCID-Profile: Yi-Fang Wang
Zitieren
APA
Wang, Y. F., Liu, F., Price, M. J., & Riegel, A. C. (2026). Impact of tumor motion on target delineation and dose calculation accuracy using rapid‐acquisition HyperSight CBCT in online adaptive radiotherapy. Journal of Applied Clinical Medical Physics, 27(1). https://doi.org/10.1002/acm2.70428
Vancouver
Wang YF, Liu F, Price MJ, Riegel AC. Impact of tumor motion on target delineation and dose calculation accuracy using rapid‐acquisition HyperSight CBCT in online adaptive radiotherapy. Journal of Applied Clinical Medical Physics. 2026;27(1). doi:10.1002/acm2.70428
BibTeX
@article{Wang2026,
author = {Wang, Yi-Fang and Liu, Fan and Price, Michael J and Riegel, Adam C},
title = {Impact of tumor motion on target delineation and dose calculation accuracy using rapid‐acquisition HyperSight CBCT in online adaptive radiotherapy},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {1},
doi = {10.1002/acm2.70428}
}
EUVIMED