Comparison of two dosimetric systems for quality assurance in a clinical treatment scenario for brain radiotherapy: Diode array and polymer gel dosimetry

Angeliki Ntouli, Georgios Kalaitzakis, John Stratakis, Kostas Perisinakis, Stefanos Kachris, Maria Tolia, Michalis Mazonakis, Thomas G Maris

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70272 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Angeliki Ntouli und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Medical School, University of Crete, Heraklion, Greece. Hauptthemen laut MEDLINE: Quality Assurance, Health Care, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Brain Neoplasms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Advanced radiotherapy techniques such as single‐isocenter volumetric modulated arc therapy (VMAT) and simultaneous integrated boost (SIB‐VMAT) require precise quality assurance (QA) due to their complexity and sensitivity to geometric and dosimetric uncertainties, especially for multi‐target configurations.

Purpose

To evaluate and compare two dosimetric systems—a diode array and a polymer gel dosimeter—for quality assurance in single‐isocenter multi‐target VMAT and SIB‐VMAT plans through 3D gamma index and statistical agreement analysis.

Methods

A diode array system (Delta⁴—Scandidos, Uppsala, Sweden) and a 3D printed phantom (Prime—RTsafe, Athens, Greece) embedded with in‐house polymer gel were utilized. The treatment plans for VMAT and SIB‐VMAT were created using the Monaco treatment planning system (TPS) and irradiations were performed with the Elekta Infinity linear accelerator with a 6‐MV photon beam on both Prime and Delta⁴. Analyses of the irradiated gels were performed using a 1.5T clinical MRI system. Additionally, 3D gamma indexes and Bland‐Altman analyses were conducted to evaluate the agreement between relative doses from MRI‐derived gel data and diode array's detector measurements.

Results

Diode array system achieved gamma passing rates (GPRs) >99%, while polymer gel showed >95% GPR for both irradiation plans. Bland–Altman analysis indicated minimal bias (mean difference: 0.1%) and narrow limits of agreement (−1.9% to 2.2%), confirming good consistency between the two dosimetric methods for both plans delivered.

Conclusions

Acceptable agreement between the two systems was observed. Both demonstrated complementary capabilities, making both essential tools for ensuring precision in advanced radiotherapy techniques. The polymer gel system offers more detailed insights compared to the diode array method, showing increased sensitivity in challenging cases that involve small targets at greater distances from the isocenter.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Quality Assurance, Health CarePhantoms, ImagingRadiotherapy Planning, Computer-AssistedBrain NeoplasmsRadiotherapy, Intensity-ModulatedPolymersRadiometryHumansRadiotherapy DosageGelsParticle Accelerators

Stichwörter der Autor:innen

MRIQuality AssuranceSIB – VMATVMATbrain radiotherapydiode arraypolymer gel dosimetry

Substanzen

PolymersGels

Autor:innen und Institutionen

  • Department of Medical Physics, Medical School, University of Crete, Heraklion, Greece.
  • Department of Medical Physics, University General Hospital of Heraklion, Heraklion, Greece.
  • Department of Radiation Oncology, University General Hospital of Heraklion, Heraklion, Greece.

Zitieren

APA

Ntouli, A., Kalaitzakis, G., Stratakis, J., Perisinakis, K., Kachris, S., Tolia, M., Mazonakis, M., & Maris, T. G. (2025). Comparison of two dosimetric systems for quality assurance in a clinical treatment scenario for brain radiotherapy: Diode array and polymer gel dosimetry. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70272

Vancouver

Ntouli A, Kalaitzakis G, Stratakis J, Perisinakis K, Kachris S, Tolia M, et al. Comparison of two dosimetric systems for quality assurance in a clinical treatment scenario for brain radiotherapy: Diode array and polymer gel dosimetry. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70272

BibTeX
@article{Ntouli2025,
  author = {Ntouli, Angeliki and Kalaitzakis, Georgios and Stratakis, John and Perisinakis, Kostas and Kachris, Stefanos and Tolia, Maria and Mazonakis, Michalis and Maris, Thomas G},
  title = {Comparison of two dosimetric systems for quality assurance in a clinical treatment scenario for brain radiotherapy: Diode array and polymer gel dosimetry},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70272}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche