Quality assurance of single isocentre multiple target stereotactic radiosurgery: Findings from long‐term off‐axis Winston‐Lutz testing and machine performance checks
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70275 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Joel Poder und 8 weiteren Autor:innen. Beteiligte Einrichtung: St George Hospital Cancer Care Centre, Kogarah, NSW, Australia. Hauptthemen laut MEDLINE: Radiosurgery, Quality Assurance, Health Care, Particle Accelerators, Phantoms, Imaging. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Single‐isocenter multiple‐target stereotactic radiosurgery (SIMT SRS) has emerged as an efficient treatment for multiple brain metastases. However, this technique demands exceptional geometric accuracy, particularly off‐axis, to ensure optimal dose delivery while sparing healthy tissue. Traditional quality assurance (QA) methods require adaptation for SIMT SRS, highlighting the need for robust testing protocols.
Purpose
This study aimed to evaluate the long‐term off‐axis targeting accuracy of three Varian TrueBeam linear accelerators using a dedicated off‐axis Winston–Lutz test (OAWLT) and to investigate correlations with routine varian machine performance check (MPC) results. Materials & Methods Weekly OAWLT measurements were performed over a 6‐month period on three TrueBeam linacs using the Sun Nuclear StereoPHAN phantom and MultiMet‐WL cube. The test delivered 6 MV flattening filter‐free beams to multiple off‐axis targets via eight beam orientations, simulating clinical SIMT SRS workflows. Concurrently, daily MPC assessments evaluated geometric parameters including isocenter size, kV isocenter offset, beam center, and couch radiation‐induced offset.
Results
Median OAWLT errors were 0.38, 0.44, and 0.59 mm for the three linacs included in this study, with maximum errors of 1.12, 1.08, and 1.54 mm. Notably, off‐axis errors increased with target off‐axis distance, with the worst performance observed at 7 cm off‐axis.
Conclusion
Routine OAWLT is an effective and sensitive QA tool for monitoring off‐axis targeting accuracy in SIMT SRS. Despite the convenience of daily MPC, it cannot substitute for a dedicated OAWLT. The study supports implementing weekly OAWLT in clinical practice to ensure high‐quality, consistent treatment delivery for SIMT SRS treatments.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- St George Hospital Cancer Care Centre, Kogarah, NSW, Australia.
- Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia.
- Institute of Medical Physics, University of Sydney, Camperdown, NSW, Australia.
- St. George and Sutherland Clinical School, University of New South Wales, Kogarah, NSW, Australia.
ORCID-Profile: Joel Poder, Johnson Yuen
Zitieren
APA
Poder, J., Chambers, R., Porter, H., Brown, R., Koprivec, D., Yuen, J., Quezeda, R., Nourbehesht, L., & Howie, A. (2025). Quality assurance of single isocentre multiple target stereotactic radiosurgery: Findings from long‐term off‐axis Winston‐Lutz testing and machine performance checks. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70275
Vancouver
Poder J, Chambers R, Porter H, Brown R, Koprivec D, Yuen J, et al. Quality assurance of single isocentre multiple target stereotactic radiosurgery: Findings from long‐term off‐axis Winston‐Lutz testing and machine performance checks. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70275
BibTeX
@article{Poder2025,
author = {Poder, Joel and Chambers, Robert and Porter, Harry and Brown, Ryan and Koprivec, Dylan and Yuen, Johnson and Quezeda, Rebecca and Nourbehesht, Lisa and Howie, Andrew},
title = {Quality assurance of single isocentre multiple target stereotactic radiosurgery: Findings from long‐term off‐axis Winston‐Lutz testing and machine performance checks},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70275}
}
EUVIMED