Influence of dose calculation algorithms on the helical diode array using volumetric‐modulated arc therapy for small targets

Tomohiro Ono iD, Hideaki Hirashima iD, Takanori Adachi iD, Hiraku Iramina iD, Takahiro Fujimoto, Megumi Uto, Mitsuhiro Nakamura, Takashi Mizowaki iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14307 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Tomohiro Ono und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology and Image-Applied Therapy, Kyoto University Graduate School of Medicine, Kyoto, Japan. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Radiotherapy Dosage, Algorithms, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Scientific Research from the Ministry of Education, Culture, Sports, Science, and Technology of Japan.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

For patient‐specific quality assurance (PSQA) for small targets, the dose resolution can change depending on the characteristics of the dose calculation algorithms.

Purpose

This study aimed to evaluate the influence of the dose calculation algorithms Acuros XB (AXB), anisotropic analytical algorithm (AAA), photon Monte Carlo (pMC), and collapsed cone (CC) on a helical diode array using volumetric‐modulated arc therapy (VMAT) for small targets. Materials and methods ArcCHECK detectors were inserted with a physical depth of 2.9 cm from the surface. To evaluate the influence of the dose calculation algorithms for small targets, rectangular fields of 2×100, 5×100, 10×100, 20×100, 50×100, and 100×100 mm 2 were irradiated and measured using ArcCHECK with TrueBeam STx. A total of 20 VMAT plans for small targets, including the clinical sites of 19 brain metastases and one spine, were also evaluated. The gamma passing rates (GPRs) were evaluated for the rectangular fields and the 20 VMAT plans using AXB, AAA, pMC, and CC.

Results

For rectangular fields of 2×100 and 5×100 mm 2 , the GPR at 3%/2 mm of AXB was < 50% because AXB resulted in a coarser dose resolution with narrow beams. For field sizes > 10×100 mm2, the GPR at 3%/2 mm was > 88.1% and comparable for all dose calculation algorithms. For the 20 VMAT plans, the GPRs at 3%/2 mm were 79.1 ± 15.7%, 93.2 ± 5.8%, 94.9 ± 4.1%, and 94.5 ± 4.1% for AXB, AAA, pMC, and CC, respectively.

Conclusion

The behavior of the dose distribution on the helical diode array differed depending on the dose calculation algorithm for small targets. Measurements using ArcCHECK for VMAT with small targets can have lower GPRs owing to the coarse dose resolution of AXB around the detector area.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy, Intensity-ModulatedRadiotherapy DosageAlgorithmsRadiotherapy Planning, Computer-AssistedMonte Carlo MethodOrgans at RiskHumansPhantoms, ImagingBrain NeoplasmsQuality Assurance, Health CareSpinal Neoplasms

Stichwörter der Autor:innen

Acuros XBGPRVMATcoarser dose resolutionsmall‐field irradiation

Autor:innen und Institutionen

  • Department of Radiation Oncology and Image-Applied Therapy, Kyoto University Graduate School of Medicine, Kyoto, Japan.
  • Division of Clinical Radiology Service, Kyoto University Hospital, Kyoto, Japan.
  • Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, Kyoto, Japan.

ORCID-Profile: Tomohiro Ono, Hideaki Hirashima, Takanori Adachi, Hiraku Iramina, Takashi Mizowaki

Förderung

  • Scientific Research from the Ministry of Education, Culture, Sports, Science, and Technology of Japan

Zitieren

APA

Ono, T., Hirashima, H., Adachi, T., Iramina, H., Fujimoto, T., Uto, M., Nakamura, M., & Mizowaki, T. (2024). Influence of dose calculation algorithms on the helical diode array using volumetric‐modulated arc therapy for small targets. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14307

Vancouver

Ono T, Hirashima H, Adachi T, Iramina H, Fujimoto T, Uto M, et al. Influence of dose calculation algorithms on the helical diode array using volumetric‐modulated arc therapy for small targets. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14307

BibTeX
@article{Ono2024,
  author = {Ono, Tomohiro and Hirashima, Hideaki and Adachi, Takanori and Iramina, Hiraku and Fujimoto, Takahiro and Uto, Megumi and Nakamura, Mitsuhiro and Mizowaki, Takashi},
  title = {Influence of dose calculation algorithms on the helical diode array using volumetric‐modulated arc therapy for small targets},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14307}
}
Kostenlos starten →