Influence of dose calculation algorithms on the helical diode array using volumetric‐modulated arc therapy for small targets
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14307 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Tomohiro Ono und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology and Image-Applied Therapy, Kyoto University Graduate School of Medicine, Kyoto, Japan. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Radiotherapy Dosage, Algorithms, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Scientific Research from the Ministry of Education, Culture, Sports, Science, and Technology of Japan.
Abstract (Originalsprache)
Background
For patient‐specific quality assurance (PSQA) for small targets, the dose resolution can change depending on the characteristics of the dose calculation algorithms.
Purpose
This study aimed to evaluate the influence of the dose calculation algorithms Acuros XB (AXB), anisotropic analytical algorithm (AAA), photon Monte Carlo (pMC), and collapsed cone (CC) on a helical diode array using volumetric‐modulated arc therapy (VMAT) for small targets. Materials and methods ArcCHECK detectors were inserted with a physical depth of 2.9 cm from the surface. To evaluate the influence of the dose calculation algorithms for small targets, rectangular fields of 2×100, 5×100, 10×100, 20×100, 50×100, and 100×100 mm 2 were irradiated and measured using ArcCHECK with TrueBeam STx. A total of 20 VMAT plans for small targets, including the clinical sites of 19 brain metastases and one spine, were also evaluated. The gamma passing rates (GPRs) were evaluated for the rectangular fields and the 20 VMAT plans using AXB, AAA, pMC, and CC.
Results
For rectangular fields of 2×100 and 5×100 mm 2 , the GPR at 3%/2 mm of AXB was < 50% because AXB resulted in a coarser dose resolution with narrow beams. For field sizes > 10×100 mm2, the GPR at 3%/2 mm was > 88.1% and comparable for all dose calculation algorithms. For the 20 VMAT plans, the GPRs at 3%/2 mm were 79.1 ± 15.7%, 93.2 ± 5.8%, 94.9 ± 4.1%, and 94.5 ± 4.1% for AXB, AAA, pMC, and CC, respectively.
Conclusion
The behavior of the dose distribution on the helical diode array differed depending on the dose calculation algorithm for small targets. Measurements using ArcCHECK for VMAT with small targets can have lower GPRs owing to the coarse dose resolution of AXB around the detector area.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology and Image-Applied Therapy, Kyoto University Graduate School of Medicine, Kyoto, Japan.
- Division of Clinical Radiology Service, Kyoto University Hospital, Kyoto, Japan.
- Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
ORCID-Profile: Tomohiro Ono, Hideaki Hirashima, Takanori Adachi, Hiraku Iramina, Takashi Mizowaki
Förderung
- Scientific Research from the Ministry of Education, Culture, Sports, Science, and Technology of Japan
Zitieren
APA
Ono, T., Hirashima, H., Adachi, T., Iramina, H., Fujimoto, T., Uto, M., Nakamura, M., & Mizowaki, T. (2024). Influence of dose calculation algorithms on the helical diode array using volumetric‐modulated arc therapy for small targets. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14307
Vancouver
Ono T, Hirashima H, Adachi T, Iramina H, Fujimoto T, Uto M, et al. Influence of dose calculation algorithms on the helical diode array using volumetric‐modulated arc therapy for small targets. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14307
BibTeX
@article{Ono2024,
author = {Ono, Tomohiro and Hirashima, Hideaki and Adachi, Takanori and Iramina, Hiraku and Fujimoto, Takahiro and Uto, Megumi and Nakamura, Mitsuhiro and Mizowaki, Takashi},
title = {Influence of dose calculation algorithms on the helical diode array using volumetric‐modulated arc therapy for small targets},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14307}
}
EUVIMED