Evaluation of a Monte Carlo based EPID system for patient‐specific IMRT and VMAT quality assurance
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70178 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Wuji Sun und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology & Therapy, The First Hospital of Jilin University, Changchun, China. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Monte Carlo Method, Radiotherapy Planning, Computer-Assisted, Quality Assurance, Health Care. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Jilin Scientific and Technological Development Program, Liaoning Provincial Science and Technology Program Joint Program, Jilin Provincial Scientific and Technological Development Program.
Abstract (Originalsprache)
Purpose
This study aims to evaluate the performance of a new electronic portal imaging device (EPID) system with a Monte Carlo (MC) based algorithm for patient‐specific quality assurance (PSQA), ensuring its reliability and effectiveness in treatment verification efficiency and precision.
Methods
Sixteen patients with various tumor sites were divided evenly into two groups for dynamic intensity‐modulated radiation therapy and volumetric modulated arc therapy verification. Measurements were performed on a UIH uRT‐linac 506c linear accelerator with a Varex Imaging XRD 1642 EPID. The performance of the EPID was assessed for sensitivity to errors in monitor units, collimator angle, field offset, field size, and multi‐leaf collimator functionality. Treatment plans were modified for each error and verified to determine the capability of the system to detect perturbations from the planned dose distribution, which was quantitatively analyzed using γ index analysis (2%/2 mm, 10% low‐dose threshold). Additionally, the ArcCHECK phantom was employed to validate the EPID system for PSQA and induced error detection. Pearson's correlation coefficient was employed to assess correlations between γ passing rates and induced errors.
Results
The EPID system showed high accuracy in dose linearity and error detection, with γ passing rates consistently above 95% for original plans. Sensitivity tests indicated strong correlations between induced errors and γ passing rates, confirming the capability of this system to detect subtle dosimetric discrepancies and supporting its application for PSQA. Both EPID and ArcCHECK demonstrated comparable γ passing rates in PSQA of standardized plans and treatment plans, with EPID showing higher or comparable sensitivity to tested errors.
Conclusion
The EPID Plan QA system showed promising accuracy and sensitivity in detecting errors, proving to be a reliable tool for PSQA. Its seamless integration into clinical workflows is expected to enhance treatment verification efficiency and ensure precise radiotherapy delivery.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology & Therapy, The First Hospital of Jilin University, Changchun, China.
- United Imaging Healthcare Co., Ltd., Shanghai, China.
- Department of Radiation Oncology, First Affiliated Hospital of Anhui Medical University, Hefei, China.
- Department of Radiation Oncology, The First Hospital of China Medical University, Shenyang, China.
- Jilin Provincial Key Laboratory of Radiation Oncology & Therapy, Department of Radiation Oncology & Therapy, The First Hospital of Jilin University, Changchun, China.
- NHC Key Laboratory of Radiobiology, School of Public Health, Jilin University, Changchun, China.
ORCID-Profile: Wuji Sun, Huidong Wang
Förderung
- Jilin Scientific and Technological Development Program
- Liaoning Provincial Science and Technology Program Joint Program
- Jilin Provincial Scientific and Technological Development Program
Zitieren
APA
Sun, W., Ge, C., Shi, Y., He, X., Xiao, L., Liu, Y., Wang, Z., Ji, T., & Wang, H. (2025). Evaluation of a Monte Carlo based EPID system for patient‐specific IMRT and VMAT quality assurance. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70178
Vancouver
Sun W, Ge C, Shi Y, He X, Xiao L, Liu Y, et al. Evaluation of a Monte Carlo based EPID system for patient‐specific IMRT and VMAT quality assurance. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70178
BibTeX
@article{Sun2025,
author = {Sun, Wuji and Ge, Chao and Shi, Yinghua and He, Xiaohe and Xiao, Li and Liu, Yanfang and Wang, Zhi and Ji, Tianlong and Wang, Huidong},
title = {Evaluation of a Monte Carlo based EPID system for patient‐specific IMRT and VMAT quality assurance},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70178}
}
EUVIMED