Initial three‐ and half‐year machine performance assessment of a superficial radiation therapy unit
Journal of Applied Clinical Medical Physics · 2025-10 · 26 (10)
DOI: 10.1002/acm2.70258 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 21.09.2025. Verfasst von Yongsook C Lee und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Miami Cancer Institute, Baptist Health South Florida, Miami, Florida, USA. Hauptthemen laut MEDLINE: Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted, Radiometry, Particle Accelerators. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
This work reports on results of periodic machine quality assurance (QA) tests for our superficial radiation therapy (SRT) unit (SRT‐100 Vision) performed for the first three‐ and half‐ year.
Methods
Results of machine warm‐up, dosimetry, mechanical, safety and imaging QA tests for our SRT‐100 Vision unit were reviewed and analyzed. Dosimetry tests included output constancy, backup timer, timer accuracy, beam quality (half value layer [HVL]), applicator factors (AFs), absolute output calibration, output constancy with varying x‐ray tube head rotation, output reproducibility, output linearity, timer and end‐effect error, percent depth dose (PDD) verification and congruence between radiation field and applicator size. Mechanical tests encompassed unit stability as well as applicator integrity and indicators. Safety tests consisted of visual and audio monitors, beam‐on indicator, interlocks, and energy/filter indicators. Imaging tests for ultrasound covered functional check, spatial integrity, and image quality check. The QA tests were performed for three kilovoltage (kV) x‐ray energies (50, 70, and 100 kV p ) and six applicators (1.5, 2.0, 2.5, 3.0, 4.0, and 5.0 cm in diameter).
Results
Daily mandatory machine warm‐up was successfully completed on each day of treatment. The results of all dosimetry tests for three energies were within the recommended tolerance. Daily, and monthly outputs and annual absolute outputs were < ± 3.0%, < ± 2.0%, and < ± 1.3%, respectively. Backup timer displayed timer set ± 0.0%. Timer accuracy was < ± 1.0 sec. HVLs and AFs were < ± 1.1% and < ± 0.3%, respectively. Outputs for different head angles were < ± 1.0% and output reproducibility was within 0.1%. Output linearity was < ± 0.6% and end‐effect errors were < ± 0.0006 min. PDDs were < ± 2.2%. Congruence was < ± 0.5 mm. Monthly mechanical tests confirmed mechanical robustness of our unit. Daily safety features were verified to be functional each day. Monthly imaging tests for ultrasound showed good functionality, image quality, and spatial integrity meeting the set tolerance.
Conclusions
The QA results demonstrated high performance and stability of our SRT‐100 Vision unit over the first three and a half years of operation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Miami Cancer Institute, Baptist Health South Florida, Miami, Florida, USA.
- Department of Radiation Oncology, Herbert Wertheim College of Medicine, Florida International University, Miami, Florida, USA.
ORCID-Profile: D Jay Wieczorek
Zitieren
APA
Lee, Y. C., Romaguera, W., Davis, S. D., Wieczorek, D. J., Chaswal, V., Tolakanahalli, R., Mehta, M. P., Kalman, N. S., & Gutierrez, A. N. (2025). Initial three‐ and half‐year machine performance assessment of a superficial radiation therapy unit. Journal of Applied Clinical Medical Physics, 26(10). https://doi.org/10.1002/acm2.70258
Vancouver
Lee YC, Romaguera W, Davis SD, Wieczorek DJ, Chaswal V, Tolakanahalli R, et al. Initial three‐ and half‐year machine performance assessment of a superficial radiation therapy unit. Journal of Applied Clinical Medical Physics. 2025;26(10). doi:10.1002/acm2.70258
BibTeX
@article{Lee2025,
author = {Lee, Yongsook C and Romaguera, William and Davis, Stephen D and Wieczorek, D Jay and Chaswal, Vibha and Tolakanahalli, Ranjini and Mehta, Minesh P and Kalman, Noah S and Gutierrez, Alonso N},
title = {Initial three‐ and half‐year machine performance assessment of a superficial radiation therapy unit},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
volume = {26},
number = {10},
doi = {10.1002/acm2.70258}
}
EUVIMED