Long‐term stability evaluation of user‐derived electrometer calibration factors ( k elec, user ) for radiotherapy dosimetry
Journal of Applied Clinical Medical Physics · 2026-10 · 27 (10)
DOI: 10.1002/acm2.70840 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 08.10.2026. Verfasst von Hayato Tsuno und 8 weiteren Autor:innen. Beteiligte Einrichtung: Gunma Prefectural College of Health Sciences, Maebashi, Gunma, JAPAN. Hauptthemen laut MEDLINE: Radiometry, Particle Accelerators, Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
User‐derived electrometer calibration factors ( k elec, user ) have been proposed as a practical approach for electrometer quality assurance in radiotherapy dosimetry. However, the contribution of long‐term stability to the uncertainty of k elec, user has not been quantitatively evaluated. This study aimed to assess the long‐term stability of k elec, user over a 28‐month period and to validate the method using different signal input techniques.
Methods
A RAMTEC Solo electrometer was used as the reference electrometer, and three electrometers (UNIDOS webline , RAMTEC Duo, and Super MAX) were evaluated as electrometers under test. Electrometer sensitivity ratios were measured using three signal input methods: A LINAC‐based method, a battery‐powered current source (SCG002), and a servo‐controlled current source (EMF521R). User‐derived calibration factors were calculated from the sensitivity ratios and the calibration coefficient of the reference electrometer. Long‐term stability was evaluated from repeated measurements over 28 months using linear regression analysis, and the associated uncertainty component was estimated according to the Guide to the Expression of Uncertainty in Measurement (GUM).
Results
The coefficients of variation for repeated measurements were ≤ 0.071% for all methods. The k elec, user values obtained using the LINAC method, SCG002, and EMF521R agreed within 0.1%, with a maximum relative difference of 0.06%. Over the 28‐month observation period, the largest estimated drift was 1.84 × 10 − 4 , corresponding to a Type B standard uncertainty of 1.06 × 10 − 4 (1.06 × 10 − 2 %). The combined standard uncertainty of k elec, user remained 0.20%, and the experimentally determined long‐term stability component had a negligible effect on the overall uncertainty.
Conclusions
The influence of signal input method on k elec, user was minimal, and consistent results were obtained using LINAC‐based and current‐source‐based measurements. The uncertainty contribution from long‐term stability over 28 months was negligible compared with the reported uncertainty of electrometer calibration factors. These findings support the use of periodic in‐house sensitivity comparison measurements as a complementary quality assurance tool for monitoring electrometer performance between external calibrations. These findings extend previous studies by providing quantitative evidence of the relative sensitivity stability of user‐derived electrometer calibration factors and thereby support their application for routine in‐house quality assurance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Gunma Prefectural College of Health Sciences, Maebashi, Gunma, JAPAN.
- Department of Radiation Oncology, Japanese Red Cross Maebashi Hospital, Maebashi, Gunma, JAPAN.
- Graduate School of Radiological Technology, Gunma Prefectural College of Health Sciences, Maebashi, Gunma, JAPAN.
- Department of Radiology, National Hospital Organization Sendai Medical Center, Sendai, Miyagi, JAPAN.
- Department of Radiation Oncology, Cancer Institute Hospital, Japanese Foundation for Cancer Research, Tokyo, JAPAN.
- Department of Radiology, JR Tokyo General Hospital, Tokyo, JAPAN.
ORCID-Profile: Hayato Tsuno, Koji Sasaki, Fumiyasu Matsubayashi
Zitieren
APA
Tsuno, H., Sasaki, K., Sasaki, R., Matsubayashi, F., Nishi, K., Ushikawa, T., Goto, D., Kosugi, R., & Kawashima, Y. (2026). Long‐term stability evaluation of user‐derived electrometer calibration factors ( k elec, user ) for radiotherapy dosimetry. Journal of Applied Clinical Medical Physics, 27(10). https://doi.org/10.1002/acm2.70840
Vancouver
Tsuno H, Sasaki K, Sasaki R, Matsubayashi F, Nishi K, Ushikawa T, et al. Long‐term stability evaluation of user‐derived electrometer calibration factors ( k elec, user ) for radiotherapy dosimetry. Journal of Applied Clinical Medical Physics. 2026;27(10). doi:10.1002/acm2.70840
BibTeX
@article{Tsuno2026,
author = {Tsuno, Hayato and Sasaki, Koji and Sasaki, Ruan and Matsubayashi, Fumiyasu and Nishi, Kohei and Ushikawa, Tae and Goto, Daisaku and Kosugi, Ririko and Kawashima, Yasuhiro},
title = {Long‐term stability evaluation of user‐derived electrometer calibration factors ( k elec, user ) for radiotherapy dosimetry},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {10},
doi = {10.1002/acm2.70840}
}
EUVIMED