EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

kV reference dosimetry in Australia and New Zealand: Survey results and trends

Madelaine Tyler, Mitchell Duncan iD, Joanne McNamara

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14458 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Madelaine Tyler und 2 weiteren Autor:innen. Beteiligte Einrichtung: Shoalhaven Cancer Care Centre, Nowra, New South Wales, Australia. Hauptthemen laut MEDLINE: Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted, Radiometry. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To assess the number of radiotherapy kilovoltage (kV) units in service, their clinical utilization, and methodology and equipment used for absorbed dose determination across Australia and New Zealand.

Methods

A survey was sent to 61 Australian and New Zealand radiotherapy providers in the second half of 2023.

Results

Fifty‐seven responses were received, with 43 departments having kV units and providing beam quality data for 185 therapeutic kV beams 20–300 kVp. Percentage depth dose curves were compared between five clinical beams with 100 kVp and 2.13–6.28 mm Aluminum half value layers (HVLs), demonstrating large differences that can occur between beams with the same kVp. Eighteen departments provided clinical utilization data for their kV units, with a total of 4458 treatment courses and their corresponding kVp reported. All departments complied with national and international recommendations with respect to the equipment used for reference dosimetry of kV beams; 77% of ionization chambers used for absorbed dose determination were of Farmer‐type, with the remaining 23% being plane parallel soft x‐ray chambers. Methods of derivation of air‐kerma calibration factors varied, with 73% of respondents using a draft document disseminated by the Australian Primary Standards laboratory, 23% using HVL alone, and 6% using other methods.

Conclusions

The results of this survey provide a snapshot of kilovoltage radiation therapy use and the number of kV units across Australia and New Zealand. This data can be used as a point of reference for future investigations into clinical utilization and reference dosimetry methods across Australia and New Zealand or for comparisons with other countries, facilitating standardization of reference dosimetry practice for kilovoltage units.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy DosageRadiotherapy Planning, Computer-AssistedRadiometryHumansNew ZealandAustraliaSurveys and QuestionnairesNeoplasmsParticle AcceleratorsCalibration

Stichwörter der Autor:innen

dosimetrykilovoltageradiotherapy

Autor:innen und Institutionen

  • Shoalhaven Cancer Care Centre, Nowra, New South Wales, Australia.
  • Townsville Cancer Centre, Townsville, Queensland, Australia.

ORCID-Profile: Mitchell Duncan

Zitieren

APA

Tyler, M., Duncan, M., & McNamara, J. (2024). kV reference dosimetry in Australia and New Zealand: Survey results and trends. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14458

Vancouver

Tyler M, Duncan M, McNamara J. kV reference dosimetry in Australia and New Zealand: Survey results and trends. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14458

BibTeX
@article{Tyler2024,
  author = {Tyler, Madelaine and Duncan, Mitchell and McNamara, Joanne},
  title = {kV reference dosimetry in Australia and New Zealand: Survey results and trends},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14458}
}
Kostenlos starten →