Implementation of optically simulated luminescent dosimeter for quality control of gamma ray dose of an accelerator‐based neutron source

Naonori Hu iD, Taiki Nakamura, Ryusuke Kataura, Keita Suga, Tetsuya Mukawa, Kazuhiko Akita, Akinori Sasaki, Mai Nojiri iD, Nishiki Matsubayashi, Takushi Takata, Hiroki Tanaka, Keiji Nihei und 1 weitere

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14493 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Naonori Hu und 12 weiteren Autor:innen. Beteiligte Einrichtung: Osaka Medical and Pharmaceutical University, Kansai BNCT Medical Center, Osaka, Japan. Hauptthemen laut MEDLINE: Gamma Rays, Boron Neutron Capture Therapy, Particle Accelerators, Neutrons. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Neutron beams utilized for performing BNCT are composed of a mixture of neutrons and gamma rays. Although much of the dose delivered to the cancer cells comes from the high LET particles produced by the boron neutron capture reaction, the dose delivered to the healthy tissues from unwanted gamma rays cannot be ignored. With the increase in the number of accelerators for BNCT, a detector system that is capable of measuring gamma ray dose in a mixed neutron/gamma irradiation field is crucial. Currently, BeO TLDs encased in quartz glass are used to measure gamma ray dose in a BNCT irradiation field. However, this type of TLD is no longer commercially available. A replacement dosimetry system is required to perform the recommended ongoing quality assurance of gamma ray measurement for a clinical BNCT system.

Purpose

The purpose of this study is to evaluate the characteristics of a BeO OSLD detector system under a mixed neutron and gamma ray irradiation field and to assess the suitability of the system for routine quality assurance measurements of an accelerator‐based BNCT facility.

Methods

The myOSLD system by RadPro International GmbH was evaluated using the accelerator‐based neutron source designed for clinical BNCT (NeuCure BNCT system). The readout constancy, linearity, dose rate effect, and fading effect of the OSLD were evaluated. Free‐in‐air and water phantom measurements were performed and compared with the TLD results and Monte Carlo simulation results. The PHITS Monte Carlo code was used for this study.

Results

The readout constancy was found to be stable over a month‐long period and similar to the TLD results. The OSLD readout signal was found to be linear, with a high coefficient of determination ( R 2 ≥ 0.999) up to a proton charge of 3.6 C. There was no significant signal fading or dose rate dependency. The central axis depth dose and off‐axis dose profile measurements agreed with both the TLD and Monte Carlo simulation results, within one standard deviation.

Conclusion

The myOSLD system was characterized using an accelerator system designed for clinical BNCT. The experimental measurements confirmed the OSLD achieved similar, if not superior to, the currently utilized dosimetry system for routine QA of an accelerator‐based BNCT system. The OSLD system would be a suitable replacement for the current TLD system for performing routine QA of gamma ray dose measurement in a BNCT irradiation field.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Gamma RaysBoron Neutron Capture TherapyParticle AcceleratorsNeutronsPhantoms, ImagingRadiotherapy DosageQuality ControlMonte Carlo MethodHumansRadiotherapy Planning, Computer-AssistedNeoplasmsRadiation DosimetersQuality Assurance, Health CareRadiometry

Stichwörter der Autor:innen

BNCTMonte Carlo simulationOSLDQAaccelerator neutron source

Autor:innen und Institutionen

  • Osaka Medical and Pharmaceutical University, Kansai BNCT Medical Center, Osaka, Japan.
  • Institute for Integrated Radiation and Nuclear Science, Kyoto University, Osaka, Japan.
  • Sumitomo Heavy Industries, Ltd., Tokyo, Japan.
  • Graduate School of Engineering, Kyoto University, Kyoto, Japan.
  • Osaka Medical and Pharmaceutical University Hospital, Department of Radiation Oncology, Osaka, Japan.

ORCID-Profile: Naonori Hu, Mai Nojiri

Zitieren

APA

Hu, N., Nakamura, T., Kataura, R., Suga, K., Mukawa, T., Akita, K., Sasaki, A., Nojiri, M., Matsubayashi, N., Takata, T., Tanaka, H., Nihei, K., & Ono, K. (2024). Implementation of optically simulated luminescent dosimeter for quality control of gamma ray dose of an accelerator‐based neutron source. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14493

Vancouver

Hu N, Nakamura T, Kataura R, Suga K, Mukawa T, Akita K, et al. Implementation of optically simulated luminescent dosimeter for quality control of gamma ray dose of an accelerator‐based neutron source. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14493

BibTeX
@article{Hu2024,
  author = {Hu, Naonori and Nakamura, Taiki and Kataura, Ryusuke and Suga, Keita and Mukawa, Tetsuya and Akita, Kazuhiko and Sasaki, Akinori and Nojiri, Mai and Matsubayashi, Nishiki and Takata, Takushi and Tanaka, Hiroki and Nihei, Keiji and Ono, Koji},
  title = {Implementation of optically simulated luminescent dosimeter for quality control of gamma ray dose of an accelerator‐based neutron source},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14493}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche