Clinical comparison of four commercial MRI phantoms on a 0.35T MR‐linac using a standard clinical TRUFI sequence at a single institution
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70247 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Mateb Al Khalifa und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Particle Accelerators, Magnetic Resonance Imaging, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Real‑time magnetic resonance–guided radiation therapy (MRgRT) integrates MRI with a linear accelerator (Linac) for gating and adaptive radiotherapy, which requires robust image‑quality assurance over a large field of view (FOV). Specialized phantoms capable of accommodating this extensive FOV are therefore essential. This study compares the performance of four commercial MRI phantoms on a 0.35 T MR‑Linac.
Methods
Four phantoms: the IBA QUASAR Insight Phantom, the IBA QUASAR MRID 3D Geometric Distortion Phantom, the MagPhan Phantom, and the Sun Nuclear Large Field MRI Distortion Phantom, were evaluated on a 0.35 T MR‐Linac using the TRUFI clinical sequence. Mean distortions were measured at gantry angles from 0° to 360° in 30° increments. Clinical imaging parameters (signal‐to‐noise ratio, uniformity, resolution via modulation transfer function, laser alignment, and twist) were also assessed where applicable.
Results
The QUASAR MRID 3D Phantom enabled separate quantification of B 0 and gradient‐induced distortions but showed a notable vertical offset, which did not affect comparisons with other phantoms. Both the MagPhan and QUASAR Insight phantoms provided mean distortion measurements as well as additional imaging parameters (e.g., signal‐to‐noise ratio, uniformity, resolution). The Sun Nuclear Phantom exhibited higher overall distortion values.
Conclusion
All four phantoms successfully measured distortion at multiple gantry angles on a 0.35 T MR‐Linac. The QUASAR MRID 3D Phantom uniquely differentiated B 0 from gradient distortions but required an offset adjustment. The MagPhan and QUASAR Insight phantoms offered comprehensive imaging metrics, while the Sun Nuclear Phantom exhibited relatively higher distortions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA.
- Department of Radiation Medicine, Northwell Health, Lake Success, New York, USA.
ORCID-Profile: Mateb Al Khalifa, Haya Aljuaid, Siyong Kim, William Y Song
Zitieren
APA
Khalifa, M. A., Ma, T., Aljuaid, H., Kim, S., & Song, W. Y. (2025). Clinical comparison of four commercial MRI phantoms on a 0.35T MR‐linac using a standard clinical TRUFI sequence at a single institution. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70247
Vancouver
Khalifa MA, Ma T, Aljuaid H, Kim S, Song WY. Clinical comparison of four commercial MRI phantoms on a 0.35T MR‐linac using a standard clinical TRUFI sequence at a single institution. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70247
BibTeX
@article{Khalifa2025,
author = {Khalifa, Mateb Al and Ma, Tianjun and Aljuaid, Haya and Kim, Siyong and Song, William Y},
title = {Clinical comparison of four commercial MRI phantoms on a 0.35T MR‐linac using a standard clinical TRUFI sequence at a single institution},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70247}
}
EUVIMED