Acceptance testing of a 0.35 T MR‐Linac: procedures, QA baseline, and system limitations

Mateb Al Khalifa iD, Tianjun Ma, Haya Aljuaid iD, Siyong Kim iD, William Y Song iD

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70488 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Mateb Al Khalifa und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Phantoms, Imaging, Particle Accelerators, Quality Assurance, Health Care. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study describes and evaluates the acceptance procedure for a ViewRay (VR) MRIdian 0.35T MR‐Linac, emphasizing key challenges, limitations, and recommendations to enhance clinical performance and accuracy.

Methods

A comprehensive acceptance test was conducted at a single institution, following the manufacturer's protocols and aligned with established acceptance guidelines. Specific tools and phantoms were used to assess three primary components: mechanical, dosimetric, and Magnetic Resonance Imaging (MRI).

Results

Overall, the test outcomes satisfied the manufacturer's specifications. However, certain issues were identified: high couch attenuation at specific gantry angles (leading to their exclusion from treatment), variations in magnetic field homogeneity at different gantry angles, and discrepancies between TPS calculations and measurements for field output factors smaller than 0.83 cm × 0.83 cm.

Conclusion

This work provides a detailed account of the acceptance testing procedure and establishes a QA baseline for 0.35T MR‐Linac systems. In doing so, it also identifies key system limitations, such as couch attenuation, magnetic field inhomogeneity, and small‐field output discrepancies, underscoring the need for careful gantry angle selection, field homogeneity optimization, and meticulous validation of very small fields.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingPhantoms, ImagingParticle AcceleratorsQuality Assurance, Health CareRadiotherapy Planning, Computer-AssistedImage Processing, Computer-AssistedHumansRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

MRgRTMR‐LinacQAacceptance test

Autor:innen und Institutionen

  • Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA.
  • Department of Radiation Medicine, Northwell Health, Lake Success, New York, USA.

ORCID-Profile: Mateb Al Khalifa, Haya Aljuaid, Siyong Kim, William Y Song

Zitieren

APA

Khalifa, M. A., Ma, T., Aljuaid, H., Kim, S., & Song, W. Y. (2026). Acceptance testing of a 0.35 T MR‐Linac: procedures, QA baseline, and system limitations. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70488

Vancouver

Khalifa MA, Ma T, Aljuaid H, Kim S, Song WY. Acceptance testing of a 0.35 T MR‐Linac: procedures, QA baseline, and system limitations. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70488

BibTeX
@article{Khalifa2026,
  author = {Khalifa, Mateb Al and Ma, Tianjun and Aljuaid, Haya and Kim, Siyong and Song, William Y},
  title = {Acceptance testing of a 0.35 T MR‐Linac: procedures, QA baseline, and system limitations},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70488}
}
Kostenlos starten →