Acceptance testing of a 0.35 T MR‐Linac: procedures, QA baseline, and system limitations
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70488 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Mateb Al Khalifa und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Phantoms, Imaging, Particle Accelerators, Quality Assurance, Health Care. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
This study describes and evaluates the acceptance procedure for a ViewRay (VR) MRIdian 0.35T MR‐Linac, emphasizing key challenges, limitations, and recommendations to enhance clinical performance and accuracy.
Methods
A comprehensive acceptance test was conducted at a single institution, following the manufacturer's protocols and aligned with established acceptance guidelines. Specific tools and phantoms were used to assess three primary components: mechanical, dosimetric, and Magnetic Resonance Imaging (MRI).
Results
Overall, the test outcomes satisfied the manufacturer's specifications. However, certain issues were identified: high couch attenuation at specific gantry angles (leading to their exclusion from treatment), variations in magnetic field homogeneity at different gantry angles, and discrepancies between TPS calculations and measurements for field output factors smaller than 0.83 cm × 0.83 cm.
Conclusion
This work provides a detailed account of the acceptance testing procedure and establishes a QA baseline for 0.35T MR‐Linac systems. In doing so, it also identifies key system limitations, such as couch attenuation, magnetic field inhomogeneity, and small‐field output discrepancies, underscoring the need for careful gantry angle selection, field homogeneity optimization, and meticulous validation of very small fields.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA.
- Department of Radiation Medicine, Northwell Health, Lake Success, New York, USA.
ORCID-Profile: Mateb Al Khalifa, Haya Aljuaid, Siyong Kim, William Y Song
Zitieren
APA
Khalifa, M. A., Ma, T., Aljuaid, H., Kim, S., & Song, W. Y. (2026). Acceptance testing of a 0.35 T MR‐Linac: procedures, QA baseline, and system limitations. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70488
Vancouver
Khalifa MA, Ma T, Aljuaid H, Kim S, Song WY. Acceptance testing of a 0.35 T MR‐Linac: procedures, QA baseline, and system limitations. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70488
BibTeX
@article{Khalifa2026,
author = {Khalifa, Mateb Al and Ma, Tianjun and Aljuaid, Haya and Kim, Siyong and Song, William Y},
title = {Acceptance testing of a 0.35 T MR‐Linac: procedures, QA baseline, and system limitations},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70488}
}
EUVIMED