EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Hounsfield Unit characterization and dose calculation on a C‐arm linac with novel on‐board cone‐beam computed tomography feature and advanced reconstruction algorithms

Kenneth W Gregg iD, Theodore Arsenault iD, Atefeh Rezaei, Rojano Kashani, Lauren Henke, Alex T Price

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70145 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Kenneth W Gregg und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University Hospitals Seidman Cancer Center, Cleveland, Ohio, USA. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Particle Accelerators. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Current cone‐beam computed tomography (CBCT) on‐board c‐arm linear accelerators (linacs) lack CT number precision sufficient for dose calculation due to increased scatter from the cone geometry. This investigation evaluated CT number and dose calculation accuracy in‐phantom on a novel on‐board CBCT unit with potential for improved dose calculation accuracy.

Methods

Eight head and eight body configurations of an electron density phantom using 16 materials were acquired with a clinical CT‐simulator(CT‐sim) and novel on‐board CBCT imager. CBCT projection data was reconstructed using conventional Feldkamp–Davis–Kress(FDK) and patient scatter‐corrected iterative Acuros CTS with metal artifact reduction (Acuros‐CTS‐iCBCT‐MAR) to create a robust CT number to physical density curve. CT‐sim and CBCT images of anthropomorphic head and thorax phantoms were acquired, and three treatment plans were generated per phantom. All CBCT images were registered to the CT‐sim of the phantoms, and the dose was recalculated on the CBCT images. 3D gamma analysis was performed (10% dose threshold, local, 1% dose difference, 1 or 2 mm distance to agreement), and dose–volume histogram (DVH) metrics were reported for target coverage and organ sparing.

Results

CT numbers for materials ≤1.08 g/cc showed high agreement between CT‐sim and CBCT acquisitions. CT number precision improved with Acuros‐CTS‐iCBCT‐MAR compared to FDK in all materials. High agreement between CT‐sim and CBCT reconstructed with Acuros‐CTS‐iCBCT‐MAR was observed in 3D gamma analysis showing 93.8% voxels passing in the worst case for a spine stereotactic body radiotherapy (SBRT) plan at 1%/1 mm. Maximum deviation in target coverage was −3.3% PTV‐D98% in the lung SBRT plan among the novel reconstructed CBCT images. Plan comparison using FDK reconstructions yielded similar or worse agreement in 3D gamma analysis, with 69.7% voxels passing in the worst case for a spine SBRT treatment plan. Maximum target coverage deviation was −11.7% PTV‐D98% among the FDK‐reconstructed CBCT.

Conclusions

Novel CBCT solutions on c‐arm linacs show promise for enabling direct‐to‐unit or offline adaptive dose calculation, potentially increasing versatility and efficiency of patient‐centered care.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyPhantoms, ImagingRadiotherapy Planning, Computer-AssistedParticle AcceleratorsAlgorithmsImage Processing, Computer-AssistedHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedHead

Stichwörter der Autor:innen

Hounsfield Unit characterizationdose calculationfast cone‐beam CTinhomogeneity correction

Autor:innen und Institutionen

  • Department of Radiation Oncology, University Hospitals Seidman Cancer Center, Cleveland, Ohio, USA.
  • Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA.
  • School of Medicine, Case Western Reserve University, Cleveland, Ohio, USA.

ORCID-Profile: Kenneth W Gregg, Theodore Arsenault

Zitieren

APA

Gregg, K. W., Arsenault, T., Rezaei, A., Kashani, R., Henke, L., & Price, A. T. (2025). Hounsfield Unit characterization and dose calculation on a C‐arm linac with novel on‐board cone‐beam computed tomography feature and advanced reconstruction algorithms. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70145

Vancouver

Gregg KW, Arsenault T, Rezaei A, Kashani R, Henke L, Price AT. Hounsfield Unit characterization and dose calculation on a C‐arm linac with novel on‐board cone‐beam computed tomography feature and advanced reconstruction algorithms. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70145

BibTeX
@article{Gregg2025,
  author = {Gregg, Kenneth W and Arsenault, Theodore and Rezaei, Atefeh and Kashani, Rojano and Henke, Lauren and Price, Alex T},
  title = {Hounsfield Unit characterization and dose calculation on a C‐arm linac with novel on‐board cone‐beam computed tomography feature and advanced reconstruction algorithms},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70145}
}
Kostenlos starten →