A streamlined workflow for comprehensive 4DCT imaging QA with automated analysis of spatial integrity and image quality
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70168 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Yunjie Yang und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA. Hauptthemen laut MEDLINE: Four-Dimensional Computed Tomography, Phantoms, Imaging, Workflow, Image Processing, Computer-Assisted. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.
Abstract (Originalsprache)
Purpose
A comprehensive 4DCT QA program that includes the assessment of spatial integrity and image quality of 4D phantom scans can be resource‐intensive, especially because the analysis burden scales with the number of motion traces used for QA. This work presents a streamlined and scalable workflow, enabled by the use of a widely available phantom and an automated analysis tool.
Methods
For 4DCT imaging QA, the Catphan was placed on the QUASAR motion platform, driven with sinusoidal motion traces of various amplitudes and frequencies. The acquired image sets were automatically analyzed using a newly designed 4DCT analysis module in TotalQA. The metrics of interest for 4DCT QA were analyzed for image sets of all motion bins and compared to the stationary reference image set. Three broad categories of imaging tests for 4DCT QA were defined: spatial integrity, HU constancy, and image quality. The sensitometry plugs in the Catphan were used as motion surrogates for evaluating spatial integrity, including the dimensions of the plugs and the observed motion amplitude in the scans. HU values of the sensitometry plugs in different phases were evaluated for consistency. For image quality, spatial resolution, low contrast resolution, and image noise were evaluated.
Results
Compared to the stationary reference CT, spatial integrity was within ±1 mm and the HU values were within two HU in the 4DCT for all phases. The spatial resolution was consistent while slightly higher noise was observed in the 4DCT images. The automated analysis in TotalQA was completed in approximately 20 min per motion trace, improving efficiency by more than 80% compared to the manual workflow.
Conclusions
A streamlined 4DCT imaging QA workflow with automated analysis offers a robust assessment of imaging quality and motion accuracy in 4DCT scans, enabling an efficient and consistent QA process in large healthcare networks.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
- Image Owl, Inc., Greenwich, New York, USA.
ORCID-Profile: Yunjie Yang, Donghoon Lee, Matthew Whitaker, Wendy Harris, Jeho Jeong, Shih-Chi Lin, Ellen Yorke, Grace Tang
Förderung
- NCI NIH HHS
Zitieren
APA
Yang, Y., Lee, D., Whitaker, M., Harris, W., Jeong, J., Lin, S. C., Yorke, E., Santanam, L., & Tang, G. (2025). A streamlined workflow for comprehensive 4DCT imaging QA with automated analysis of spatial integrity and image quality. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70168
Vancouver
Yang Y, Lee D, Whitaker M, Harris W, Jeong J, Lin SC, et al. A streamlined workflow for comprehensive 4DCT imaging QA with automated analysis of spatial integrity and image quality. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70168
BibTeX
@article{Yang2025,
author = {Yang, Yunjie and Lee, Donghoon and Whitaker, Matthew and Harris, Wendy and Jeong, Jeho and Lin, Shih-Chi and Yorke, Ellen and Santanam, Lakshmi and Tang, Grace},
title = {A streamlined workflow for comprehensive 4DCT imaging QA with automated analysis of spatial integrity and image quality},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70168}
}
EUVIMED