Automated Analysis for MR Coil QA
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70624 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Bhudatt Paliwal und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Human Oncology, University of Wisconsin Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Phantoms, Imaging, Quality Assurance, Health Care, Image Processing, Computer-Assisted. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Phantom‐based quality assurance (QA) of magnetic resonance imaging (MRI) coils is a standard practice for assessing signal‐to‐noise ratio (SNR), image uniformity, and magnetic field homogeneity. Manual slice selection and region‐of‐interest (ROI) placements introduce operator‐dependent variability and workflow inefficiencies.
Purpose
To develop a unified, fully automated desktop‐based workflow for MRI coil QA that eliminates manual slice selection and ROI placement. Materials and methods A Python desktop app was built to automate MRI coil QA workflows for body, torso, and head‐and‐neck coils. It takes DICOM image sets and raw free‐induction decay data and performs slice selection, phantom detection, and fixed‐geometry ROI placement automatically to calculate standard QA metrices: SNR, percent image uniformity, and frequency‐domain magnetic field homogeneity. Thus, obtained results are directly saved in structured reports and plots.
Results
The automated workflow carried out all implemented QA tasks, without manual ROI placement, and consistently produced QA metrices for body, torso, head‐and‐neck, weekly QA, and magnetic field homogeneity analyses. The standardization of the outputs facilitated an objective evaluation of the coils and channel performance by eliminating the subjective bias.
Conclusions
This study presents an automated MRI quality assurance workflow that operates through standardized testing methods and delivers consistent results while decreasing the impact of human error in standard clinical quality control procedures.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Human Oncology, University of Wisconsin Madison, Madison, Wisconsin, USA.
Zitieren
APA
Paliwal, B., Bal, A., Kodali, L., Tewatia, D., & Tanumihardjo, I. (2026). Automated Analysis for MR Coil QA. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70624
Vancouver
Paliwal B, Bal A, Kodali L, Tewatia D, Tanumihardjo I. Automated Analysis for MR Coil QA. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70624
BibTeX
@article{Paliwal2026,
author = {Paliwal, Bhudatt and Bal, Aviral and Kodali, Lokesh and Tewatia, Dinesh and Tanumihardjo, Idarto},
title = {Automated Analysis for MR Coil QA},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70624}
}
EUVIMED