Enhancing proton therapy quality assurance with custom‐designed Octopoint phantom and Gafchromic film

Xueyan Tang, Chris J Beltran, Keith M Furutani iD, Graham S Gilson, Jon Gustafson, Michael G Herman, Shima Ito, Jed E Johnson iD, Jon J Kruse, Kenneth M Long, Daniel W Mundy, Nicholas B Remmes und 4 weitere

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70156 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Xueyan Tang und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Proton Therapy, Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Proton therapy offers precise tumor targeting while minimizing damage to surrounding healthy tissue, making it especially valuable for treating tumors near critical organs and in pediatric patients. However, its success depends on accurate beam delivery, requiring rigorous quality assurance (QA) to maintain treatment precision and effectiveness.

Purpose

This study aims to standardize the measurement of key aspects of proton therapy delivery, such as gantry and couch isocentricity, spot position accuracy, and spot size consistency. The goal is to enhance the delivery accuracy of proton therapy across various clinical settings, improving patient outcomes.

Methods

The QA framework uses the Octopoint phantom, Gafchromic film, and spot position monitor (SPM) log data to evaluate proton beam isocentricity, spot position, and spot size. The Octopoint phantom, made from acrylic, was used with Gafchromic films at various gantry and couch angles to measure isocentricity. MATLAB tools were used to analyze spot positions, and SPM logs provided verification. Sensitivity tests were conducted to assess the system's response to intentional shifts and errors, ensuring alignment with clinical QA standards.

Results

The Octopoint phantom's stepwise two‐dimensional (2D) fitting process, validated against film‐measured dose profiles, accurately identified beam and ball bearing (BB) centroid locations. Isocentricity tests conducted over 8 months across four gantries demonstrated consistent beam‐to‐BB radial offsets. The phantom showed excellent repeatability, with a maximum standard deviation of 0.1 mm across various couch‐gantry combinations. Sensitivity testing across all axes revealed a strong alignment between intended shifts and measured values. Over the course of a year, film measurements tracked spot position and size consistency, with deviations remaining within acceptable clinical limits. Comparisons with SPM data further confirmed the system's reliability in maintaining beam accuracy across different gantry angles and energy levels.

Conclusions

This study presents a reliable QA framework for ensuring precision in proton therapy delivery. By combining the Octopoint phantom, Gafchromic film analysis, and SPM log file evaluation, we effectively measured isocentricity, spot position accuracy, and spot size stability. The framework demonstrated adaptability across various clinical QA tasks, enhancing the accuracy and safety of proton therapy treatments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingProton TherapyQuality Assurance, Health CareRadiotherapy Planning, Computer-AssistedFilm DosimetryNeoplasmsHumansRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

Isocentricityquality assurancespot scanning proton therapy

Autor:innen und Institutionen

  • Department of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA.

ORCID-Profile: Keith M Furutani, Jed E Johnson, Ali M Tasson, Thomas J Whitaker

Zitieren

APA

Tang, X., Beltran, C. J., Furutani, K. M., Gilson, G. S., Gustafson, J., Herman, M. G., Ito, S., Johnson, J. E., Kruse, J. J., Long, K. M., Mundy, D. W., Remmes, N. B., Tasson, A. M., Whitaker, T. J., Matysiak, W., & Tryggestad, E. J. (2025). Enhancing proton therapy quality assurance with custom‐designed Octopoint phantom and Gafchromic film. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70156

Vancouver

Tang X, Beltran CJ, Furutani KM, Gilson GS, Gustafson J, Herman MG, et al. Enhancing proton therapy quality assurance with custom‐designed Octopoint phantom and Gafchromic film. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70156

BibTeX
@article{Tang2025,
  author = {Tang, Xueyan and Beltran, Chris J and Furutani, Keith M and Gilson, Graham S and Gustafson, Jon and Herman, Michael G and Ito, Shima and Johnson, Jed E and Kruse, Jon J and Long, Kenneth M and Mundy, Daniel W and Remmes, Nicholas B and Tasson, Ali M and Whitaker, Thomas J and Matysiak, Witold and Tryggestad, Erik J},
  title = {Enhancing proton therapy quality assurance with custom‐designed Octopoint phantom and Gafchromic film},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70156}
}
Kostenlos starten →