Optimization of image reconstruction technique for respiratory-gated lung stereotactic body radiotherapy treatment planning using four-dimensional CT: a phantom study

Kenji Yasue, Hiraku Fuse, Minori Takaoka, Shin Miyakawa, Norikazu Koori, Masato Takahashi, Kazuya Shinoda, Hideaki Ikoma, Tatsuya Fujisaki, Shinji Abe

Radiological Physics and Technology · 2025

DOI: 10.1007/s12194-024-00850-6 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiological Physics and Technology, 2025. Verfasst von Kenji Yasue und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, 4669-2, Ami, Inashiki, Ibaraki, 300-0394, Japan. yasueke@ipu.…. Hauptthemen laut MEDLINE: Radiosurgery, Four-Dimensional Computed Tomography, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 18 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Patient respiration is characterized by respiratory parameters, such as cycle, amplitude, and baseline drift. In treatment planning using four-dimensional computed tomography (4DCT) images, the target dose may be affected by variations in image reconstruction techniques and respiratory parameters. This study aimed to optimize 4DCT image reconstruction techniques for the treatment planning of lung stereotactic body radiotherapy (SBRT) based on respiratory parameters using respiratory motion phantom. We quantified respiratory parameters using 30 respiratory motion datasets. The 4DCT images were acquired, and the phase- and amplitude-based reconstruction images (RI) were created. The target dose was calculated based on these reconstructed images. Statistical analysis was performed using Pearson's correlation coefficient (r) to determine the relationship between respiratory parameters and target dose in each reconstructed technique and respiratory region. In the inhalation region of phase-based RI, r of the target dose and baseline drift was -0.52. In particular, the target dose was significantly reduced for respiratory parameters with a baseline drift of 0.8 mm/s and above. No other respiratory parameters or respiratory regions were significantly correlated with target dose in phase-based RI. In amplitude-based RI, there were no significant differences in the correlation between all respiratory parameters and target dose in the exhalation or inhalation regions. These results showed that the target dose of the amplitude-based RI did not depend on changes in respiratory parameters or respiratory regions, compared to the phase-based RI. However, it is possible to guarantee the target dose by considering respiratory parameters during the inhalation region of the phase-based RI.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryFour-Dimensional Computed TomographyPhantoms, ImagingRadiotherapy Planning, Computer-AssistedRespirationLung NeoplasmsImage Processing, Computer-AssistedLungRespiratory-Gated Imaging TechniquesHumansRadiotherapy Dosage

Stichwörter der Autor:innen

4DCTAmplitude-based RILung cancerPhase-based RIRespiratory gating techniqueStereotactic body radiation therapy

Autor:innen und Institutionen

  • Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, 4669-2, Ami, Inashiki, Ibaraki, 300-0394, Japan. yasueke@ipu.ac.jp.
  • Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, 4669-2, Ami, Inashiki, Ibaraki, 300-0394, Japan.
  • Master's Program in Medical Sciences, Graduate School of Comprehensive Human Sciences, University of Tsukuba, 1-1-1, Tennoudai, Tsukuba, Ibaraki, 305-8577, Japan.
  • Department of Radiation Technology, Ibaraki Prefectural Central Hospital, 6528, Koibuchi, Kasama, Ibaraki, 309-1793, Japan.

Zitieren

APA

Yasue, K., Fuse, H., Takaoka, M., Miyakawa, S., Koori, N., Takahashi, M., Shinoda, K., Ikoma, H., Fujisaki, T., & Abe, S. (2025). Optimization of image reconstruction technique for respiratory-gated lung stereotactic body radiotherapy treatment planning using four-dimensional CT: a phantom study. Radiological Physics and Technology. https://doi.org/10.1007/s12194-024-00850-6

Vancouver

Yasue K, Fuse H, Takaoka M, Miyakawa S, Koori N, Takahashi M, et al. Optimization of image reconstruction technique for respiratory-gated lung stereotactic body radiotherapy treatment planning using four-dimensional CT: a phantom study. Radiological Physics and Technology. 2025. doi:10.1007/s12194-024-00850-6

BibTeX
@article{Yasue2025,
  author = {Yasue, Kenji and Fuse, Hiraku and Takaoka, Minori and Miyakawa, Shin and Koori, Norikazu and Takahashi, Masato and Shinoda, Kazuya and Ikoma, Hideaki and Fujisaki, Tatsuya and Abe, Shinji},
  title = {Optimization of image reconstruction technique for respiratory-gated lung stereotactic body radiotherapy treatment planning using four-dimensional CT: a phantom study},
  journal = {Radiological Physics and Technology},
  year = {2025},
  doi = {10.1007/s12194-024-00850-6}
}