EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY 4.0

Efficient EPID‐based quality assurance of beam time delay for respiratory‐gated radiotherapy with validation on Catalyst™ and AlignRT™ systems

Kaining Yao, Meijiao Wang, Yi Du iD, Jiacheng Liu, Qingying Wang, Ruoxi Wang, Hao Wu, Haizhen Yue

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14376 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Kaining Yao und 7 weiteren Autor:innen. Beteiligte Einrichtung: Key laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Canc…. Hauptthemen laut MEDLINE: Radiotherapy Dosage, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Science Foundation of Peking University Cancer Hospital, Clinical Research Project of Wu Jieping Medical Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To propose a straightforward and time‐efficient quality assurance (QA) approach of beam time delay for respiratory‐gated radiotherapy and validate the proposed method on typical respiratory gating systems, Catalyst™ and AlignRT™.

Methods

The QA apparatus was composed of a motion platform and a Winston‐Lutz cube phantom (WL3) embedded with metal balls. The apparatus was first scanned in CT‐Sim and two types of QA plans specific for beam on and beam off time delay, respectively, were designed. Static reference images and motion testing images of the WL3 cube were acquired with EPID. By comparing the position differences of the embedded metal balls in the motion and reference images, beam time delays were determined. The proposed approach was validated on three linacs with either Catalyst™ or AlignRT™ respiratory gating systems. To investigate the impact of energy and dose rate on beam time delay, a range of QA plans with Eclipse (V15.7) were devised with varying energy and dose rates.

Results

For all energies, the beam on time delays in AlignRT™ V6.3.226, AlignRT™ V7.1.1, and Catalyst™ were 92.13 5.79 ms, 123.11 6.44 ms, and 303.44 4.28 ms, respectively. The beam off time delays in AlignRT™ V6.3.226, AlignRT™ V7.1.1, and Catalyst™ were 121.87 1.34 ms, 119.33 0.75 ms, and 97.69 2.02 ms, respectively. Furthermore, the beam on delays decreased slightly as dose rates increased for all gating systems, whereas the beam off delays remained unaffected.

Conclusions

The validation results demonstrate the proposed QA approach of beam time delay for respiratory‐gated radiotherapy was both reproducible and time‐efficient to practice for institutions to customize accordingly.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy DosagePhantoms, ImagingRadiotherapy Planning, Computer-AssistedRadiotherapy, Intensity-ModulatedQuality Assurance, Health CareHumansParticle AcceleratorsRespirationRespiratory-Gated Imaging TechniquesNeoplasmsTime Factors

Stichwörter der Autor:innen

dose ratequality assurancerespiratory gatingrespiratory‐gated radiotherapytime delay

Autor:innen und Institutionen

  • Key laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing, China.
  • Institute of Medical Technology, Peking University Health Science Center, Beijing, China.

ORCID-Profile: Yi Du

Förderung

  • National Natural Science Foundation of China
  • Science Foundation of Peking University Cancer Hospital
  • Clinical Research Project of Wu Jieping Medical Foundation
  • National Key Research and Development Program of China

Zitieren

APA

Yao, K., Wang, M., Du, Y., Liu, J., Wang, Q., Wang, R., Wu, H., & Yue, H. (2024). Efficient EPID‐based quality assurance of beam time delay for respiratory‐gated radiotherapy with validation on Catalyst™ and AlignRT™ systems. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14376

Vancouver

Yao K, Wang M, Du Y, Liu J, Wang Q, Wang R, et al. Efficient EPID‐based quality assurance of beam time delay for respiratory‐gated radiotherapy with validation on Catalyst™ and AlignRT™ systems. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14376

BibTeX
@article{Yao2024,
  author = {Yao, Kaining and Wang, Meijiao and Du, Yi and Liu, Jiacheng and Wang, Qingying and Wang, Ruoxi and Wu, Hao and Yue, Haizhen},
  title = {Efficient EPID‐based quality assurance of beam time delay for respiratory‐gated radiotherapy with validation on Catalyst™ and AlignRT™ systems},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14376}
}
Kostenlos starten →