Validierungsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Heath Davis und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Virginia, Charlottesville, Virginia, USA. Hauptthemen laut MEDLINE: Beryllium, Optically Stimulated Luminescence Dosimetry, Radiotherapy Planning, Computer-Assisted, Radiometry. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
In radiation oncology, inter‐fractional dosimetry using optically stimulated luminescent detectors (OSLDs) ensures accurate plan delivery and patient safety. RadPro International GmbH's myOLSchip system, featuring a beryllium oxide (BeO) OSL dosimeter, reader, and eraser, was characterized and calibrated with a Varian Truebeam for in‐vivo dosimetry following AAPM TG‐191 guidelines. The BeO detectors demonstrated good dose linearity and repeatability across multiple exposures and erasure cycles, aligning with the manufacturer's stated accuracy and precision.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiation Oncology, University of Virginia, Charlottesville, Virginia, USA.
Zitieren
APA
Davis, H., Siebers, J. V., Wijesooriya, K., & Mistro, M. (2025). Clinical validation of myOSLchip: A beryllium oxide optically stimulated luminescent dosimeter (OSLD) system in radiotherapy dosimetry. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70094
Vancouver
Davis H, Siebers JV, Wijesooriya K, Mistro M. Clinical validation of myOSLchip: A beryllium oxide optically stimulated luminescent dosimeter (OSLD) system in radiotherapy dosimetry. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70094
BibTeX
@article{Davis2025,
author = {Davis, Heath and Siebers, Jeffrey V and Wijesooriya, Krishni and Mistro, Matt},
title = {Clinical validation of myOSLchip: A beryllium oxide optically stimulated luminescent dosimeter (OSLD) system in radiotherapy dosimetry},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70094}
}
EUVIMED