EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Developing reference plans for evaluating global clinical trials credentialing and PSQA systems

Fre'Etta M D Brooks iD, Mohammad Hussein, Jessica Lye, Christopher L Nelson, Nakamura Mitsuhiro, Mallory C Glenn, Patricia Diez iD, Rushil Patel iD, Maddison Shaw, Ileana Silvestre Patallo iD, Miriam Barry, Catharine H Clark und 2 weitere

Journal of Applied Clinical Medical Physics · 2025-07 · 26 (7)

DOI: 10.1002/acm2.70113 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 29.05.2025. Verfasst von Fre'Etta M D Brooks und 13 weiteren Autor:innen. Beteiligte Einrichtung: University of Washington Medical Center, Seattle, Washington, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Credentialing, Organs at Risk. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, Public Health Service.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To develop a practical framework for creating a diverse set of validated reference plans (varying in complexity) and implement a workflow to introduce beam modeling, calibration, and delivery errors into the reference cohort to test and compare various dosimetry audit methodologies.

Methods

Sixteen IMRT and VMAT reference plans were created, using RayStation software, for four phantom geometries based on established credentialing cases from participating Global Harmonization Group (GHG) members. These reference plans were first validated in a multi‐ion‐chamber phantom. Nine dosimetric errors (perturbations) were introduced into the plans by modifying beam model and/or delivery parameters (MLC‐offset, MLC‐transmission, leaf‐tip‐width, PDD, beam calibration, and MLC‐position) based on documented community distributions of errors; this produced 144 plans. The dose impact on the clinical target volume (CTV) and organs at risk (OARs) was determined, and a range of classifications was developed to determine if the perturbed plan should pass or should fail an audit.

Results

Introducing errors into the reference plans impacted each plan differently. Dose perturbations ranged from 10% in the mean dose to the CTV and 30% in the near maximum dose to OAR (D0.03). The 144 plans included clear “acceptable” and “unacceptable” scenarios, with significant changes in dose (relative to baseline reference values), as well as near pass/fail threshold results. Plan complexity was found to have a strong impact on dose deviation, and the mean MLC Gap metric was found to best capture this relationship.

Conclusion

This study presented a framework to develop a set of reference plans and perturbations that can be used to assess and compare various audit and PSQA methodologies. The GHG has developed this framework as part of our ongoing work to test the comparability of their audit systems; this framework supports our work of aligning international dosimetry audits across the globe.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedRadiotherapy, Intensity-ModulatedCredentialingOrgans at RiskPhantoms, ImagingClinical Trials as TopicNeoplasmsQuality Assurance, Health CareHumansRadiotherapy DosageRadiometryCalibrationSoftware

Stichwörter der Autor:innen

IMRT/VMATdosimetry audit comparisoninternation dosimetry auditquality assurancesensitivity specificity evaluation

Autor:innen und Institutionen

  • University of Washington Medical Center, Seattle, Washington, USA.
  • Metrology for Medical Physics, National Physical Laboratory, Teddington, UK.
  • Olivia Newton John Cancer Wellness and Research Centre, Austin Health, Melbourne, Australia.
  • Australian Clinical Dosimetry Service, ARPANSA, Melbourne, Australia.
  • RMIT University, Melbourne, Australia.
  • Department of Radiation Physics, University of Texas MD Anderson Cancer Center, Houston, Texas, USA.

ORCID-Profile: Fre'Etta M D Brooks, Patricia Diez, Rushil Patel, Ileana Silvestre Patallo, Joerg Lehmann

Förderung

  • NCI NIH HHS
  • Public Health Service

Zitieren

APA

Brooks, F. M. D., Hussein, M., Lye, J., Nelson, C. L., Mitsuhiro, N., Glenn, M. C., Diez, P., Patel, R., Shaw, M., Patallo, I. S., Barry, M., Clark, C. H., Lehmann, J., & Kry, S. F. (2025). Developing reference plans for evaluating global clinical trials credentialing and PSQA systems. Journal of Applied Clinical Medical Physics, 26(7). https://doi.org/10.1002/acm2.70113

Vancouver

Brooks FMD, Hussein M, Lye J, Nelson CL, Mitsuhiro N, Glenn MC, et al. Developing reference plans for evaluating global clinical trials credentialing and PSQA systems. Journal of Applied Clinical Medical Physics. 2025;26(7). doi:10.1002/acm2.70113

BibTeX
@article{Brooks2025,
  author = {Brooks, Fre'Etta M D and Hussein, Mohammad and Lye, Jessica and Nelson, Christopher L and Mitsuhiro, Nakamura and Glenn, Mallory C and Diez, Patricia and Patel, Rushil and Shaw, Maddison and Patallo, Ileana Silvestre and Barry, Miriam and Clark, Catharine H and Lehmann, Joerg and Kry, Stephen F},
  title = {Developing reference plans for evaluating global clinical trials credentialing and PSQA systems},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  volume = {26},
  number = {7},
  doi = {10.1002/acm2.70113}
}
Kostenlos starten →