Characterization, commissioning, and clinical evaluation of a commercial BeO optically stimulated luminescence (OSL) system
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70057 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Joseph P Kowalski und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina, USA. Hauptthemen laut MEDLINE: Optically Stimulated Luminescence Dosimetry, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Radiometry. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
This article investigates the performance of a commercial BeO optically stimulated luminescent (OSL) dosimetry system (myOSLchip, RadPro GmbH International, Remscheid, Germany) through the application of the commissioning framework for luminescent dosimeters as described in the American Association of Physicists in Medicine Task Group 191 (AAPM TG191) report. Initial clinical experiences and dosimetric results are also presented. The following properties of the system were characterized: linearity correction factors ranged from ‐0.5% to +3% for dose levels spanning 0.1 to 20 Gy. Beam quality correction factors (relative to 6 MV) ranged from ‐4.5% (2.5FFF) to +4.5% (15MV) for photon beams and +1.9% (6 MeV) to +4.3% (20 MeV) for electron beams. An average (µ) signal loss per reading of ‐2.13% ± 0.20% was measured, however greater signal loss was observed in the first reading (µ = ‐2.6% ± 0.46%). An initial decline in individual element sensitivity relative to baseline was observed from 0–15 Gy cumulative dose (µ = ‐1.98% ± 0.55%), with negligible further deterioration from 15–32 Gy (µ = ‐2.38% ± 0.85%). Post‐irradiation, there was a transient OSL signal which faded with a half‐life of 1.8 min; this signal enhancement was +5% at 5 min post‐irradiation and +1% at 15 min relative to 24 h. Dosimeter response was not dependent on average dose rate in the range of 100–2500 MU/min. With respect to clinical testing, equal or superior performance compared with aluminum oxide OSLs (nanoDots) is shown for a range of clinical techniques and modalities including TSET, TBI, en‐face electrons, and pacemaker/out‐of‐field measurements. The feasibility of myOSLchip to serve as a primary clinical in vivo dosimetry system and direct replacement for Landauer's microStar system is demonstrated.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina, USA.
- Department of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA.
Zitieren
APA
Kowalski, J. P., Erickson, B. G., Wu, Q., Li, X., & Yoo, S. (2025). Characterization, commissioning, and clinical evaluation of a commercial BeO optically stimulated luminescence (OSL) system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70057
Vancouver
Kowalski JP, Erickson BG, Wu Q, Li X, Yoo S. Characterization, commissioning, and clinical evaluation of a commercial BeO optically stimulated luminescence (OSL) system. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70057
BibTeX
@article{Kowalski2025,
author = {Kowalski, Joseph P and Erickson, Brett G and Wu, Qiuwen and Li, Xinyi and Yoo, Sua},
title = {Characterization, commissioning, and clinical evaluation of a commercial BeO optically stimulated luminescence (OSL) system},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70057}
}
EUVIMED