Beam quality correction factors for dose measurements around 192 Ir brachytherapy sources

Zoi Thrapsanioti, Vasiliki Peppa, Costas J Hourdakis, Pantelis Karaiskos, Aristea Lekatou, Evaggelos Pantelis

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.14575 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Zoi Thrapsanioti und 5 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, Athens, Greece. Hauptthemen laut MEDLINE: Brachytherapy, Iridium Radioisotopes, Radiotherapy Dosage, Monte Carlo Method. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To provide beam quality correction factors () for detectors used in 192 Ir brachytherapy dosimetry measurements. Materials and Methods Ten detectors were studied, including the PTW 30013 and Exrading12 Farmer large cavity chambers, seven medium (0.1–0.3 cm 3 ) and small (< 0.1 cm 3 ) cavity chambers, and a synthetic microdiamond detector. The k Q,Qo correction factors were calculated at distances from 1 to 10 cm away from an 192 Ir source, using the EGSnrc Monte Carlo (MC) code. All detectors were calibrated in a 60 Co 10 × 10 cm 2 reference field provided by standard calibration laboratories. The impact of the central electrode, stem and wall on the detectors’ responses in 192 Ir photon energies was investigated. An experimental procedure was additionally applied for dose measurements around a microSelectron‐v2 192 Ir high dose rate (HDR) brachytherapy source using a motorized water phantom.

Results

Farmer chambers underestimated the dose near the source due to signal volume averaging effects, resulting in k Q,Qo values ranging from 1.177 and 1.317 at 1 cm, decreasing with distance to between 0.980 and 1.005 at 10 cm. Small cavity volume detectors should be used close to the source. The k Q,Qo for the studied small and medium cavity volume detectors were found to be close to unity (within 1.3%), showing also a small dependence on source‐to‐detector distance, except for ion chambers containing high‐Z materials in their construction. The presence of high‐Z materials caused an overresponse in these detectors, resulting in k Q,Qo values ranging from 0.950 at 1 cm to 0.729 at 10 cm away from the source. A dose rate constant of (1.114 ± 0.023)cGyh −1 U −1 was found in agreement with the literature (within 0.5%). Conclusions k Q,Qo values were calculated for dose measurements around 192 Ir brachytherapy sources. Farmer chambers should be preferred for measurements at increased distances, whereas small or medium cavity volume detectors, not containing high‐Z materials, should be used close to the source.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrachytherapyIridium RadioisotopesRadiotherapy DosageMonte Carlo MethodPhantoms, ImagingHumansRadiometryCalibrationRadiotherapy Planning, Computer-AssistedNeoplasmsPhotons

Stichwörter der Autor:innen

beam quality correction factorsbrachytherapydosimetry measurementquality assurance

Substanzen

Iridium RadioisotopesIridium-192

Autor:innen und Institutionen

  • Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, Athens, Greece.
  • Ionizing Radiation Unit, Greek Atomic Energy Commission (EEAE), Agia Paraskevi, Greece.
  • Radiotherapy department, Alexandra Hospital, Athens, Greece.

Zitieren

APA

Thrapsanioti, Z., Peppa, V., Hourdakis, C. J., Karaiskos, P., Lekatou, A., & Pantelis, E. (2025). Beam quality correction factors for dose measurements around 192 Ir brachytherapy sources. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14575

Vancouver

Thrapsanioti Z, Peppa V, Hourdakis CJ, Karaiskos P, Lekatou A, Pantelis E. Beam quality correction factors for dose measurements around 192 Ir brachytherapy sources. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.14575

BibTeX
@article{Thrapsanioti2025,
  author = {Thrapsanioti, Zoi and Peppa, Vasiliki and Hourdakis, Costas J and Karaiskos, Pantelis and Lekatou, Aristea and Pantelis, Evaggelos},
  title = {Beam quality correction factors for dose measurements around 192 Ir brachytherapy sources},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.14575}
}
Kostenlos starten →