Evaluating the use of diagnostic CT with flattening filter free beams for palliative radiotherapy: Dosimetric impact of scanner calibration variability

Madeleine L Van de Kleut, Lesley A Buckley iD, Elsayed S M Ali

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70040 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Madeleine L Van de Kleut und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Radiation Oncology and Medical Physics, The Ottawa Hospital, Ottawa, Canada. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Tomography, X-Ray Computed, Palliative Care, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Palliative radiotherapy comprises a significant portion of the radiation treatment workload. Volumetric‐modulated arc therapy (VMAT) improves dose conformity and, in conjunction with flattening filter free (FFF) delivery, can decrease treatment times, both of which are desirable in a population with a high probability of retreatment with large palliative doses per fraction. Combining FFF and VMAT delivery with planning based on previously acquired diagnostic computed tomography (CT) scans has the potential to further expedite palliative treatment. This study evaluated the dosimetric uncertainty of using FFF beams with VMAT delivery on CT images acquired from different diagnostic vendors, and between different x‐ray tube energies, in the palliative setting.

Methods

CT‐relative electron density (CT‐RED) curves were acquired for the local CT simulator at 100, 120, and 140 kVp, and for two diagnostic CT scanners at 120 kVp. Thirty palliative VMAT plans were recalculated for each CT‐RED curve, with 6 MV flat, 6 FFF, and 10 FFF beams. The doses to 95% and 2% of the PTV, the maximum point dose to the spinal canal and esophagus, and the mean dose to the kidneys were compared between recalculated plans.

Results

Comparing the dose clouds for a given fluence map calculated with CT‐RED curves from different CT scanners at 120 kVp, the mean dose difference was at most 0.3% for each DVH metric. Similar results were reported when comparing dose clouds calculated with CT‐RED curves for 100, 120, and 140 kVp on the CT simulator.

Conclusion

The results of this study confirm that diagnostic scans acquired on machines different from the CT simulator associated with the TPS, are appropriate for VMAT treatment planning in the palliative setting with FFF photon beams.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedTomography, X-Ray ComputedPalliative CareRadiotherapy, Intensity-ModulatedPhantoms, ImagingNeoplasmsHumansRadiotherapy DosageOrgans at RiskCalibrationRadiometry

Stichwörter der Autor:innen

CT‐RED curveHounsfield unitdiagnostic CTflattening filter freepalliative radiotherapyvolumetric‐modulated arc therapy

Autor:innen und Institutionen

  • Department of Radiology, Radiation Oncology and Medical Physics, The Ottawa Hospital, Ottawa, Canada.

ORCID-Profile: Lesley A Buckley

Zitieren

APA

Van de Kleut, M. L., Buckley, L. A., & Ali, E. S. M. (2025). Evaluating the use of diagnostic CT with flattening filter free beams for palliative radiotherapy: Dosimetric impact of scanner calibration variability. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70040

Vancouver

Van de Kleut ML, Buckley LA, Ali ESM. Evaluating the use of diagnostic CT with flattening filter free beams for palliative radiotherapy: Dosimetric impact of scanner calibration variability. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70040

BibTeX
@article{VandeKleut2025,
  author = {Van de Kleut, Madeleine L and Buckley, Lesley A and Ali, Elsayed S M},
  title = {Evaluating the use of diagnostic CT with flattening filter free beams for palliative radiotherapy: Dosimetric impact of scanner calibration variability},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70040}
}
Kostenlos starten →