EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Commissioning of a motion management system for a 1.5T Elekta Unity MR‐Linac: A single institution experience

Blake R Smith iD, Joel St-Aubin iD, Daniel E Hyer

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70005 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Blake R Smith und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Iowa, Iowa City, Iowa, USA. Hauptthemen laut MEDLINE: Particle Accelerators, Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging, Magnetic Resonance Imaging. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This work describes a single institution experience of commissioning a real‐time target tracking and beam control system, known as comprehensive motion management, for a 1.5 T Elekta MR‐Linac.

Methods

Anatomical tracking and radiation beam control were tested using the MRI 4D Quasar motion phantom. Multiple respiratory breathing traces were modeled across a range of realistic regular and irregular breathing patterns ranging between 10 and 18 breaths per minute. Each of the breathing traces was used to characterize the anatomical position monitoring (APM) accuracy, and beam latency, and to quantify the dosimetric impact of both parameters during a respiratory‐gated delivery using EBT3 film dosimetry. Additional commissioning tasks were performed to verify the dosimetric constancy during beam gating and to expand our existing quality assurance program.

Results

It was determined that APM correctly predicted the 3D position of a dynamically moving tracking target to within 1.5 mm for 95% of the imaging frames with no deviation exceeding 2 mm. Among the breathing traces investigated, the mean latency ranged between −21.7 and 7.9 ms with 95% of all observed latencies within 188.3 ms. No discernable differences were observed in the relative profiles or cumulative output for a gated beam relative to an ungated beam with minimal dosimetric impact observed due to system latency. Measured dose profiles for all gated scenarios retained a gamma pass rate of 97% or higher for a 3%/2 mm criteria relative to a theoretical gated dose profile without latency or tracking inaccuracies.

Conclusion

MRI‐guided target tracking and automated beam delivery control were successfully commissioned for the Elekta Unity MR‐Linac. These gating features were shown to be highly accurate with an effectively small beam latency for a range of regular and irregular respiratory breathing traces.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Particle AcceleratorsRadiotherapy Planning, Computer-AssistedPhantoms, ImagingMagnetic Resonance ImagingQuality Assurance, Health CareRespiratory-Gated Imaging TechniquesRadiotherapy, Image-GuidedHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedRespirationMovementMotionImage Processing, Computer-Assisted

Stichwörter der Autor:innen

MR‐Linaccommissioninggatingmotion managementunity

Autor:innen und Institutionen

  • Department of Radiation Oncology, University of Iowa, Iowa City, Iowa, USA.

ORCID-Profile: Blake R Smith, Joel St-Aubin

Zitieren

APA

Smith, B. R., St-Aubin, J., & Hyer, D. E. (2025). Commissioning of a motion management system for a 1.5T Elekta Unity MR‐Linac: A single institution experience. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70005

Vancouver

Smith BR, St-Aubin J, Hyer DE. Commissioning of a motion management system for a 1.5T Elekta Unity MR‐Linac: A single institution experience. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70005

BibTeX
@article{Smith2025,
  author = {Smith, Blake R and St-Aubin, Joel and Hyer, Daniel E},
  title = {Commissioning of a motion management system for a 1.5T Elekta Unity MR‐Linac: A single institution experience},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70005}
}
Kostenlos starten →