EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Portability of IMRT QA between matched linear accelerators

Brendan Barraclough, Zacariah E Labby, Sean P Frigo

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14492 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Brendan Barraclough und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina, USA. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted, Particle Accelerators. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To determine if patient‐specific IMRT quality assurance can be measured on any matched treatment delivery system (TDS) for patient treatment delivery on another.

Methods

Three VMAT plans of varying complexity were created for each available energy for head and neck, SBRT lung, and right chestwall anatomical sites. Each plan was delivered on three matched Varian TrueBeam TDSs to the same Scandidos Delta 4 Phantom+ diode array with only energy‐specific device calibrations. Dose distributions were corrected for TDS output and then compared to TPS calculations using gamma analysis. Round‐robin comparisons between measurements from each TDS were also performed using point‐by‐point dose difference, median dose difference, and the percent of point dose differences within 2% of the mean metrics.

Results

All plans had more than 95% of points passing a gamma analysis using 3%/3 mm criteria with global normalization and a 20% threshold when comparing measurements to calculations. The tightest gamma analysis criteria where a plan still passed > 95% were similar across delivery systems—within 0.5%/0.5 mm for all but three plan/energy combinations. Median dose deviations in measurement‐to‐measurement comparisons were within 0.7% and 1.0% for global and local normalization, respectively. More than 90% of the point differences were within 2%.

Conclusion

A set of plans spanning available energies and complexity levels were delivered by three matched TDSs. Comparisons to calculations and between measurements showed dose distributions delivered by each TDS using the same DICOM RT‐plan file meet tolerances much smaller than typical clinical IMRT QA criteria. This demonstrates each TDS is modeled to a similar accuracy by a common class (shared) beam model. Additionally, it demonstrates that dose distributions from one TDS show small differences in median dose to the others. This is an important validation component of the common beam model approach, allowing for operational improvements in the clinic.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy, Intensity-ModulatedRadiotherapy DosageRadiotherapy Planning, Computer-AssistedParticle AcceleratorsQuality Assurance, Health CarePhantoms, ImagingHead and Neck NeoplasmsHumansLung Neoplasms

Stichwörter der Autor:innen

IMRT QAbeam matchmatched linacspatient specific QAportability

Autor:innen und Institutionen

  • Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina, USA.
  • Department of Human Oncology, University of Wisconsin - Madison, Madison, Wisconsin, USA.

Zitieren

APA

Barraclough, B., Labby, Z. E., & Frigo, S. P. (2024). Portability of IMRT QA between matched linear accelerators. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14492

Vancouver

Barraclough B, Labby ZE, Frigo SP. Portability of IMRT QA between matched linear accelerators. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14492

BibTeX
@article{Barraclough2024,
  author = {Barraclough, Brendan and Labby, Zacariah E and Frigo, Sean P},
  title = {Portability of IMRT QA between matched linear accelerators},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14492}
}
Kostenlos starten →